ADVANCED SETUP (Continued)
6. Number Format
To set the number format to the local requirements:
•
From the
Advanced Setup
screen, press the button the
right of the Number Format parameter.
•
Use the Up and Down arrows to select the number format.
Range:
1,234.5
1.234,5
•
Press the
Save
button to save the new setting and close
the window.
•
Press the
Cancel
button to close the window without
saving the new setting.
7. Current Time
To set the current local time in 24-hour notation:
•
From the
Advanced Setup
screen, press the button to the
right of the Current Time parameter.
•
Use the Up and Down arrows to set the hours.
Range:
00 to 23
•
Use the Up and Down arrows to set the minutes.
Range:
00 to 59
•
Press the
Save
button to save the new setting and close
the window.
•
Press the
Cancel
button to close the window without
saving the new setting.
NOTE:
Making changes to the date or time during an
active patient therapy will automatically stop the therapy.
Wait until the current patient therapy has ended before
making any adjustments to this feature.
8. Date Format
•
To set the date format to the local requirements:
•
From the
Advanced Setup
screen, press the button to the
right of the Date Format parameter.
•
Use the Up and Down arrows to select the date format.
Range: MM/DD/YY
DD/MM/YY
DD/MM/YYYY
•
Press the
Save
button to save the new setting and close
the window. Press the
Cancel
button to close the window
without saving the new setting.
9. Current Date
To set the current local date:
•
From the
Advanced Setup
screen, press the button to the
right of the Current Date parameter.
•
Use the Up and Down arrows to set the Day.
Range: 01 to 31 (depending on month)
•
Use the Up and Down arrows to set the Month.
Range: 01 to 12
•
Use the Up and Down arrows to set the Year.
Range: 2018 to 2099
•
Press the
Save
button to save the new setting and close
the window.
•
Press the
Cancel
button to close the window without
saving the new setting.
NOTE:
Making changes to the date or time during an
active patient therapy will automatically stop the therapy.
Wait until the current patient therapy has ended before
making any adjustments to this feature.
10. Software Versions
The current controller and graphics software versions
installed on the device are displayed.
11. Last Calibration Date
The last calibration date of the device is displayed.
12. Next Calibration Due
The next calibration date of the device is displayed in
number of uses and hours.