Operator’s Manual
12
SYSTEM COMPONENTS
Figure IV-1.
Product illustration
Control Module
The Arctic Sun™ Temperature Management System Control
Module is the main component of the Arctic Sun™
Temperature Management System. It incorporates the system
electronics, hydraulics, software, and the touch-screen Control
Panel. The power cord, fluid delivery line, patient temperature
in cables, patient temperature out cable and fill tube connect to
the rear of the Control Module. The figure above identifies the
main features and component connection sites. Use only BD
supplied cables and accessories with Arctic Sun™
Temperature Management System Control Module. The use of
accessories, transducers or cables other than those specified
may result in increased electromagnetic compatibility (EMC)
emissions or decreased immunity of the Arctic Sun™
Temperature Management System Control Module.
Power Cord
Hospital-grade power cords are available with several inlet
connector styles to meet national/regional power requirements
and wall outlet specifications.
Fluid Delivery Line
Water flows between the ArcticGel™ pads and Control Module
through the Fluid Delivery Line.
Fill Tube
A Fill Tube is used to fill the Control Module water reservoir.
Arctic Sun Cleaning Solution
Arctic Sun Cleaning Solution is added when filling the Control
Module with Sterile Water. This is done to suppress
microorganism growth in the water reservoir and hydraulic
circuit.
For information regarding safe handling, please refer to
http://www.medivance.com/manuals
for the Cleaning Solution
SDS.
Patient Temperature Input Cables and Probes
Patient temperature control with the Arctic Sun™ Temperature
Management System requires patient temperature feedback
provided by an indwelling patient temperature probe connected
to the Control Module via the Temp In 1 and Temp In 2 inputs.
Any commercially-available Yellow Springs Instrument 400
Series (YSI 400) compatible patient temperature probe can be
connected to the Arctic Sun™ Temperature Management
System.
The Arctic Sun™ Temperature Management System
temperature cables are available with several connector styles
(e.g. Phillips, GE, Rusch, Bard, Nellcor) for compatibility with
various manufacturers’ temperature probes. Select the patient
temperature cable with the correct connector style for your
chosen temperature probe.
Figure IV-2.
Temperature input cables
A temperature cable and probe connected to the Temp In 1
connector provides the patient temperature feedback required
for automatic patient temperature control.
A second patient temperature probe and cable are
recommended to be connected into the Temp In 2 Connector.
Temp In 2 is used to provide monitoring from a second patient
Patient
Temp 1
Patient
Temp 2
Fluid
Delivery
Line
Power
Switch
Patient
Temp
Out
Serial
Number
Fill Tube
Parking
Power
Cord
Air
Filter
USB
Port
Ethernet
Port
RS232
Port