+49 8142 448 460
www.medin-medical.com
medin Medical Innovations GmbH
Adam-Geisler-Str. 1
D-82140 Olching
OP_medinCNO EN
Rev 12 Stand 06.03.2019
63
von
68
Determination of test values for electromagnetic immunity of casings against high-frequency, wireless
communication equipment
Test
frequency
MHz
Frequency
band
MHz
Radiocommu
nication
service
Modu
lation
Maximum
output
W
Distance
m
Electromag
netic
immunity
test level
V/m
385
380 to 390
TETRA 400
Pulse
modulati
on
18 Hz
1.8
0.3
27
450
430 to 470
GMRS 460,
FRS 460
FM
± 5 kHz
deviation
1 kHz sine
2
0.3
28
710
704 to 787
LTE band 13,
17
Pulse
modulati
on 217 Hz
0.2
0.3
9
745
780
810
800 to 960
GSM
800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE band 5
Pulse
modulati
on
18 Hz
2
0.3
28
870
930
1720
1700 to
1990
GSM 1800;
CDMA 1900,
GSM 1900;
DECT;
LTE band 1,
3, 4, 25;
UMTS
Pulse
modulati
on 217 Hz
2
0.3
28
1845
1970
2450
2400 to
2570
Bluetooth,
WLAN 802.11
b/g/n,
RFID 2450,
LTE band 7
Pulse
modulati
on 217 Hz
2
0.3
28
5240
5100 to
5800
WLAN 802.11
a/n
Pulse
modulati
on 217 Hz
0.2
0.3
9
5500
5785
Summary of Contents for medinCNO