Voltage
interruptions
according to
IEC 61000-4-11
0% U
T
;
250/300 cycle
0% U
T
;
250/300 cycle
Power frequency
(50/60 Hz) magnetic
field IEC 61000-4-8
30 A/m
30 A/m
The strength of power-
frequency magnetic fields
should correspond to that
of a typical commercial or
hospital environment.
Note:
U
T
is the a.c. supply voltage prior to application of the test level
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic
environment - guidance
Conducted RF
disturbances
according to IEC
61000-4-6
3 V
0.15 MHz to 80 MHz
6 V in ISM frequency
bands
a
between
0.15 MHz and
80 MHz
80% AM at 1 kHz
3 V
0.15 MHz to 80 MHz
6 V in ISM frequency
bands
a
between
0.15 MHz and 80 MHz
80% AM at 1 kHz
Portable and mobile RF
communications equipment
should be used no closer to
any part of the medinCNO
®
,
including cables, than the
recommended separation
distance of 30 cm.
Radiated RF
disturbances
according
to
IEC 61000-4-3
3 V/m
80 MHz to 2.7 GHz
80% AM at 1 kHz
3 V/m
80 MHz to 2.7 GHz
80% AM at 1 kHz
Portable and mobile RF
communications equipment
should be used no closer to
any part of the medinCNO
®
,
including cables, than the
recommended separation
distance of 30 cm.
a
The ISM bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553
MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz and 40.66 MHz to 40.70 MHz.