16
von
68
Figure 3-4: USB and RS232 cable with ferrite
a)
RS232 cable without ferrite b) RS232 cable with a ferrite that is still open whereby the cable
must be carried through the core twice, that is, wrap the ferrite core once. c)
RS232 cable with
ferrite
d)
USB cable with ferrite
(the cable must also be carried through the ferrite core twice as
in the case of RS232 cable)
3.4.3
RS232 interface
The RS232 interface can be installed optionally. The function, data and settings are identical to the USB
interface. A ferrite (Würth Elektronik - Type: 742-711-32, gray or black; REF 39-125) with a double-wound cable
must likewise be present on the RS232 cable identical to the USB cable (see chapter 3.4.2).
Attention:
The medinCNO
®
data interface must be connected using the cable, including ferrite core (REF 39-125) specified
above (see Figure 3-4) and a computer system that meets the medically relevant requirements in terms of
electrical safety and EMC, otherwise the functionality and EMC of the medinCNO
®
cannot be guaranteed.