OP_medinCNO EN
Rev 12 Stand 06.03.2019
24
von
68
4.2
Connecting the medinCNO
®
and setting up the CPAP system
Figure 4-1: Connections of the medinCNO
®
: a) Gas feed oxygen; b) Gas feed air;
c) Mains power supply (power supply unit); d) Patient gas outlet; e) CPAP pressure meter line inlet,
f) Overpressure valve, g) USB interface, h) RS232 interface (optional), i) External alarm
Gas feed connection:
The oxygen inlet "
O
2
-IN
" (Figure 4-1 -a) and air inlet "
Air-IN
" (Figure 4-1 -b) of the medinCNO
®
must be
connected to an external air and oxygen source using gas feed tubes before the device is switched on. The air
and oxygen must comply with the quality requirements laid down in chapter 3.3.
Attention:
The medicinal air and oxygen supplied to the medinCNO
®
must meet the quality requirements laid down in
chapter 3.3 as otherwise the medinCNO
®
could be damaged.
Power supply connection:
The power connection of the medinCNO
®
"
Power
" (Figure 4-1-c) must be connected to the mains by means of
the power supply unit for the device. Users are recommended to keep the medinCNO
®
connected to the mains
at all times while it is in use. The internal rechargeable battery in the medinCNO
®
allows the device to keep
working in the event of a brief interruption in mains power supply. The medinCNO
®
can run on a fully charged
battery for up to approximately three hours. The green LED of the power supply unit shows the operating
status of the power adapter.