Please note:
When selecting the push flow and sensitivity, it is important to ensure that the pressure during the push will
become greater than the upper threshold of the sensitivity selected. Otherwise the apnoea phase will not be
interrupted by the push.
Display:
When operating in Apnoea CPAP mode, the current respiration rate (
RR
- breaths >1 mbar per minute) and
number of pushes automatically generated by the medinCNO
®
per minute (
R
insp
) are shown in the orange bar
on the top of the screen in addition to the information referred to in chapter 5.2. In the particular case of
NIPPV, the current respiration rate RR is not displayed.
If the backup function is activated, this is also displayed in the top left-hand area.
5.9
SNIPPV mode
The difference between the SNIPPV mode and the Apnoea CPAP mode is that in this case, an automatic push
is triggered not only in the event of apnoea but also synchronously with inspiration.
Also, in addition to the general settings and displays, such as:
-
Flow
(chapter 5.3)
-
Oxygen concentration
FiO
2
(chapter 5.3) and start and discontinuation of the
O
2
flush
(chapter 5.5)
-
F
insp
(additional push flow) and manual
PUSH
(chapter 5.4)
-
Alarm settings (chapter 5.12)
-
Trend graphs (chapter 5.13)
there are special settings for this mode. Its overview opens by pressing button p in Figure 5-2. They can be
changed by selecting, setting the desired value using the dial and confirming this selection by pressing the dial.
Figure 5-4: SNIPPV mode settings a)
Inspiration time
; b)
Expiration pause
; c)
Apnoea time
; d)
Sensitivity
e)
Backup on/off
and f)
Backup rate
g)
the level of the
O
2
flush
. The sensitivity is displayed on the left-hand
edge of the CPAP curve as a pair of blue arrows. The blue lines serve only to clarify the meaning of sensitivity,