05.8;08.0;021;02.6;S; ;B;070;25;120;10.6; ; ; ;0.2;02.2;
07.3;08.0;021;02.6;S; ;B;070;25;120;11.2; ; ; ;0.2;02.2;
…
02.5;06.0;026;02.5;O; ; ; ; ; ; ;07;10;05.4; ; ;
04.1;06.0;036;03.7;O; ; ; ; ; ; ;14;06;04.1; ; ;
04.1;06.0;039;03.7;O; ; ; ; ; ; ;19;06;04.9; ; ;
***;2014-04-07;14:22:51;medin-cno1004;record complete;
Start of alarm recording:
-
###;
-
Date: YYYY-MM-DD;
(current date, set in the service menu)
-
Time: HH:MM:SS;
(current time, set in the service menu,
the switch to local time or summer
time must
be performed manually)
-
Device name: medin-cnoZZZZZ
(set in the service menu)
-
SW version: VXXX
(installed main software version)
-
Alarm record;
Start and alarm recording:
-
For each device start, date and time are saved as such:
start; YYYY-MM-DD;HH:MM;
-
For each new alarm, day, time, all active alarms of high, medium and low priority and the duration
(min:sec) of the alarms are saved:
alarm; DD; HH:MM;HHHH;MMMM;LLLL;for MM:SS
The alarm coding takes place analogously to the coding in the live data (for detailed information, see chapter
3.4.4.1). The date is only shown as a day. The month can be determined using the last prior start entry. The
alarm duration is indicated in minutes and seconds from the start of the alarm until its end.
Recording the set and measured values:
The trend data contain all values saved in the trend and additional values since the last time the device was
switched on. Every 20 seconds, a line is saved. If a value is not available in the current mode, it is replaced with
blank spaces.
The structure is as follows:
-
Average value of the CPAP pressure in XX.X mbar
-
Set flow in XX.X L/min + if applicable, correction flow with activated leak assist
-
Measured FiO
2
in XXX%
-
Set, additional push flow in XX.X L/min
-
Current mode: C = CPAP, A = Apnoea CPAP, S = SNIPPV, O = Oscillation, - = Standby
-
Measured respiration rate in XXX / min
-
Status of backup
(B = actively running; -= currently not actively running or generally off )
-
Push frequency R
insp
in XXX / min