+49 8142 448 460
www.medin-medical.com
medin Medical Innovations GmbH
Adam-Geisler-Str. 1
D-82140 Olching
OP_medinCNO EN
Rev 12 Stand 06.03.2019
21
von
68
Environmental conditions of the power supply unit during operation:
Temperature:
- between 0°C and 40°C
- do not use power supply unit in locations exposed to significant
temperature fluctuations.
Moisture:
- use only in dry locations
- relative humidity
≤90%, non
-condensing
- do not use in locations exposed to significant amounts of moisture or
condensate
- do not use in locations exposed to significant environmental stress
- do not use outdoors
Vibrations:
do not use in locations exposed to constant vibration
Environmental conditions for the medinCNO
®
during transport/short-term storage:
Temperature:
-20°C to 50°C (longer-term storage only at room temperature)
Relative air humidity: 20% to 80% (not condensing)
Moisture:
During transport and storage, the medinCNO
®
must be protected from
wet conditions.
Cleanliness:
The medinCNO
®
must be protected from contamination during transport
and storage.
Attention:
If the medinCNO
®
is exposed to temperatures of ≤ 15°C during transport, it must be ac
climatised at room
temperature for at least 24 hours at the location where it is to be used before being switched on and used for
the first time. If this is not done, the condensate which has formed could damage the medinCNO
®
.
The medinCNO
®
may only be transported in the original packaging, since it contains internal lithium batteries
which may only be transported under special conditions. If the medinCNO
®
contains defective batteries, these
may not be transported and must be removed in advance and disposed of in accordance with local regulations.
(See separate instructions on transporting the medinCNO
®
and the batteries it contains).
Long-term storage of the medinCNO
®
:
Temperature:
Room temperature (approximately 20°C)
Relative air humidity: 20% to 80% (not condensing)
Moisture:
During storage, the medinCNO
®
must be protected from wet conditions.
Cleanliness:
The medinCNO
®
must be protected from contamination during storage.
Please note:
Long-term storage must take place under room temperature conditions, since the batteries may become
damaged otherwise and need to be replaced sooner.