surface needs to be clean and free of clutter and large enough to accommodate all materials
required in addition to the transducer and cable. Fill one of the bowls with fresh water, one with
a mild detergent mixed with water and the last with 16 ounces of “The Works”.
cleaning:
Step 1:
Holding the cable 6 inches from the transducer, immerse the unit in the bowl
containing the mild detergent and stir for 20-30 seconds. Remove and rinse in the
bowl containing the fresh water, using the same stirring motion used in the mile
detergent. Rinse and wipe dry.
Step 2:
Holding the body of the transducer with one hand so that you are looking at the
retaining screen protecting the sensor, carefully remove the sensor nose piece by
simply unscrewing it from the sensor body.
Do not touch the sensor diaphragm
with your finger or any other object. Also, do not try to dry the inside portion
of the transducer, as you risk damaging the pressure sensor
.
Step 3:
Place the transducer in a vertical position with the pressure sensing end facing
downward in the bowl containing “The Works” solution for approximately 15-20
seconds. Rinse in the bowl containing clean water and wipe dry the external casing
only. Place the protective screen in the same solution for 15-20 seconds, rinse and
wipe dry.
Step 4
Holding the transducer in a vertical position so that you can see the face of the
pressure sensor, screw the protective nose piece back into place.