Measurement Specialties
KPSI
®
Digital Series User’s Manual
Page 24
www.Meas-Spec.com
2
nd
u is +0=Celsius or
+1=Fahrenheit.
Use
“XUT”
command to set units.
R0
R1
.
.
.
R9
Start
Continuous
Measurements
(unimplemented)
aR0!
aR1!
.
.
.
AR9!
a
Note: address return indicates
unimplemented command
V
Initiate Verify
Sequence
aV!
:
aD0!
a0013
:
a (
Service Request
response)
aefg
“e” is
executed:
+1=YES, +0=NO
“f” is +
number unexpected interrupts
“g” is +
interrupt code
Example: a +1 + 0 + 0 is normal
XE
Set Field
Calibration
Offset
aXEou!
“o” is signed
offset
.
“u” is
units
: +0=feet of
water, +1=psi.
:
aD0!
Example:
aXE-0.05+0!
(set offset to -0.05 feet)
a0011
:
a (
Service Request
response)
Service Request
resp. returned in
001 seconds.
“1” is
number values
in“
D0
” resp:
ao
“o” = signed
offset
value (psi)
XS
Recalculate
Field Calibration
Offset
aXS!
or
aXSdu!
“d” and “u” are optional
signed values. When
omitted, sensor is assumed
to be vented to
atmosphere. When
supplied, “d” is desired
reading for sensor and in
units indicated by “u.” The
sensor will make a
measurement and
recalculate the field offset
to ensure reading matches
values entered.
:
aD0!
attt1
:
a (
Service Request
response)
Service Request
resp. returned in ttt
seconds.
“1” is
number values
in“
D0
” resp:
ao
“o” is new
offset
(psi).