Measurement Specialties
KPSI
®
Digital Series User’s Manual
Page 22
www.Meas-Spec.com
Measurement
command was not
issued, the command was aborted
by sending a new command before
the measurement time expired, or an
expected
Service Request
response was not yet received.
I
Identify Sensor
aI!
a13-KPSI---500---hhh-ssssss-vvv
13 Supports
SDI Version 1.3
commands
KPSI Manufacturer
500
Model
number
hhh
Hardware revision
ssssss Serial number
vvv
Software
revision
M
or
M0
Start Non-
Concurrent
Pressure
Measurement
aM!
:
aD0!
attt2
:
a (
Service Request
response)
“2" is
number values
in”
D0
" resp:
apu
or aputu
“P” is ±
pressure
value
“u” is +
units
indicator. Use
“XUP”
command to set units. If
“
XUM +1
" command executed then ±
temperature value and units code
(tu) also appended.
M1
Start Non-
Concurrent
Factory
Pressure
Measurement
(factory
calibrated value
always in psi)
AM1!
:
aD0!
attt1
:
a (
Service Request
response)
“1" is
number values
in“
D0
" resp:
ap
“p” is ±
pressure
value (psi)
M2
Start Non-
Concurrent
Temperature
Measurement
aM2!
:
aD0!
attt2
:
a (
Service Request
response)
“2" is
number values
in“
D0
" resp:
atu
“t” is signed
temperature
“u” is +
units
: 0=Celsius,
1=Fahrenheit. Use “
XUT
command to set units.