balanSys REV – Extension Gap First: 3in1 –
57
Fig. 130
Fig. 131
Insert the Neck Stopper Curved (79.02.0750) into the
box of the femur component.
Tighten the stem screw of the component with the
Torque Wrench 3.5 with 2 clicks while counteracting
the torque with the Neck Stopper Curved (79.02.0750).
After implantation, the stem screw must
be re-tightened with the Torque Wrench
(18.410-RAL5002) with 2 clicks.
The leg should be in extension during hardening of
the bone cement.
Avoid hyperextension during curing of the bone ce-
ment.
2x Click