balanSys REV – Extension Gap First: 3in1 –
51
Fig. 111
Fig. 112
Fig. 113
The information on the Memory Board (79.02.0637)
can assist in collecting and assembling the correct
implants.
In this example, the following implants have been
chosen:
• balanSys REV Femur C left
• balanSys REV Stem 18 / 140 off. 4 uncem.
• balanSys REV Femur Augmentation C / 5 distal
medial and lateral
• balanSys REV Femur Augmentation C / 5 dorsal
Position the chosen balanSys REV Femur onto the Fe-
mur Assembly Block (79.02.0540).
Unpack the chosen balanSys REV Augmentations
and screw the implants to the distal surface of the
balanSys REV Femur.
Snap-in the augmentation screws onto the Screw
Positioner (79.02.0270). Use it to place the screws
into the correct thread and tighten the screws with
the Hexagonal Screwdriver (314.270).
Remark
One augmentation screw is packed together with
each implant.
Surgical technique
Femur implant assembly