20
– balanSys REV – Extension Gap First: 3in1
Fig. 27
Fig. 28
Fig. 29
Insert the Reference Plate (77.02.0031) in the slot of
the black Assembly Device (79.02.0271). Insert the
Test Stem Core straight (79.02.0668) or with offset
(79.02.0669) into the hole.
Remark
The orientation of the Test Stem Core is correct when
it can no longer be rotated. The laser marking is orien-
tated towards the Reference Plate.
Position the Trial Tibial Plateau (79.02.0060 to
79.02.0064) onto the Test Stem Core.
Put the Offset Alignment Disk (79.02.0287) onto the
Trial Tibial Plateau (put the open rectangle window
over the anterior nose of the Trial Tibial Plateau).
Surgical technique
Trial tibial implant assembly