40
– balanSys REV – Extension Gap First: 3in1
Fig. 81
Fig. 82
Fig. 83
The pictogram on the upper surface highlights the
exact position. The 12 o’clock position must face to-
wards anterior.
Snap-in the Trial Stem Screw (79.02.0071) onto the
Screw Positioner (79.02.0270) and insert the screw
into the hole, but do not tighten it yet!
Rotate the Offset Alignment Disk with the Trial Femur
until the Reference Plate corresponds with the prede-
termined position.