24
– balanSys REV – Extension Gap First: 3in1
Fig. 38
Fig. 39
Fig. 40
Assess the correct femur size:
• Check the size of the primary femur implant – both
in ML and AP
• And / or use the Femoral Sizing Guide (79.02.0530
to 79.02.534) to assess the correct size
• Check if the assessed femur size is compatible
with the determined tibia size (Appendix 1 – Size
Compatibility of the balanSys REV Implants)
Remark
The Medical advisors recommend using a larger size,
if in doubt.
Use the Memory Board (79.02.0637) to record the
suggested size and side (in this example: REV Femur
size C, left leg).
If necessary, drill a pilot-hole with the 8.5 mm in-
tramedullary Drill Bit (71.02.3009).
Surgical technique
Femoral preparation