![Mathys balanSys REV Surgical Techniques Download Page 56](http://html1.mh-extra.com/html/mathys/balansys-rev/balansys-rev_surgical-techniques_1740165056.webp)
56
– balanSys REV – Extension Gap First: 3in1
Fig. 127
Fig. 128
Fig. 129
The surface of the tibial plateau needs to be free of
any foreign matter (e. g. tissue fragments, bone or
cement particles) before inserting the inlay. Snap-in
the balanSys REV Inlay into the tibial plateau.
Remark
Instead of inserting the final Inlay, a suitable Trial Inlay
may be inserted during the hardening of the bone ce-
ment. After hardening, the Trial Inlay must be re-
moved. Afterwards, the correct balanSys REV Inlay
must be inserted.
Insert the stabilization screw (packed together with
the balanSys REV Inlay).
Insert the Neck Stopper Straight (79.02.0027) into the
anterior window of the tibial plateau during tighten-
ing of the stabilization screw. It will counteract against
the applied torque.
Tighten the stabilization screw with the Torque
Wrench (18.410-RAL5002) with 2 clicks.
2x Click