PERFORMANCE SPECIFICATIONS
The table below shows Arms (Accuracy Root Mean Square) values measured using the M-LNCS/LNCS Sensor with Masimo
SET Oximetry Technology in a clinical study.
M-LNCS/LNCS - Adtx/Pdtx
Range
A
rms
90-100%
1.64%
80-90%
1.07%
70-80%
1.55%
M-LNCS/LNCS - Inf/Neo
Range
A
rms
90-100%
1.85%
80-90%
1.44%
70-80%
0.89%
SaO
2
versus error (SpO
2
– SaO
2
) with linear regression fit and upper 95% and lower 95% limits of agreement