13
12
0.7
0
100
120
No
CH1 & CH2 Positive
1
0
125
120
No
CH1 & CH2 Positive
1
0
125
120
No
CH1 & CH2 Positive
0.7
0
100
120
No
CH1 & CH2 Positive
1
0
75
120
No
CH1 & CH2 Positive
0.7
0
50
120
No
CH1 & CH2 Positive
0.7
0
40
120
No
CH1 & CH2 Positive
0.7
0
30
120
No
CH1 & CH2 Positive
0.8
0
20
120
No
CH1 & CH2 Positive
1
0
15
120
No
CH1 & CH2 Positive
1.4
0
10
120
No
CH1 & CH2 Positive
1.5
0
8
120
No
CH1 & CH2 Positive
P11
Modulation
2-100Hz
260µs-150µs Continue
Rate & width
over 6 seconds
P12
Modulation
80< - >7Hz
260µs
Continue
Rate & width
over 6 seconds
Program
The pre-set parameters of the EMS 13-24
Details
programs
are as given below:
Mode
SYN/ Rate Width Ramp ON Time Off Time Timer
ALT (Hz) (µs) (Sec) (Sec)
(Sec)
(Min)
P13
ACL repair/joint
SYN 35
300
3
8
24
20
protection back muscle
P14
Spasm small
SYN 80
300
3
10
5
20
muscle
P15
Spasm
SYN 80
250
2
8
4
20
Postoperative
P16
Arthroscopy
SYN 25
200
2
6
30
15
P17
Disuse atrophy
SYN 35
300
2
5
15
30
P18
Shoulder
SYN 50
300
5
15
50
15
Subluxation
P19
Range of motion
SYN 40
250
3
6
21
30
muscle re-education of hips
P20
Muscle training
SYN 50
250
2
10
10
20
P21
Muscle training
SYN 50
250
2
14
14
20
P22
Muscle training
SYN 35
400
2
10
10
20
P23
Muscle training
ALT 50
250
2
10
10
20
P24
Muscle training
ALT 50
250
2
10
14
20
23 Timer
Adjustable, from 5 to 60 minutes minutes
and continue(C), 5 minutes each step
24 Patient
This unit can store 60 sets of operation
Compliance Meter records. Total recorded time is 999 hours.
25 Low Battery
A low battery indicator will show up when
Indicator
the battery is low.
26 Operating
Temperature:0°~40°C
Condition
Relative Humidity: 30%~75%
Atmosphere Pressure : 700Hpa~1060Hpa
27 Remark
There may be up to a +/-10% tolerance of
all parameters and +/-20% tolerance of
output amplitude & voltage.