background image

13

12

    0.7 

  0 

  100 

  120 

No 

CH1 & CH2  Positive

    1 

  0 

  125 

  120 

No 

CH1 & CH2  Positive

    1 

  0 

  125 

  120 

No 

CH1 & CH2  Positive

    0.7 

  0 

  100 

  120 

No 

CH1 & CH2  Positive

    1 

  0 

  75 

  120 

No 

CH1 & CH2  Positive

    0.7 

  0 

  50 

  120 

No 

CH1 & CH2  Positive

    0.7 

  0 

  40 

  120 

No 

CH1 & CH2  Positive

    0.7 

  0 

  30 

  120 

No 

CH1 & CH2  Positive

    0.8 

  0 

  20 

  120 

No 

CH1 & CH2  Positive

    1 

  0 

  15 

  120 

No 

CH1 & CH2  Positive

  1.4 

  0 

  10 

  120 

No 

CH1 & CH2  Positive

  1.5 

  0 

  8 

  120 

No 

CH1 & CH2  Positive

 

P11 

Modulation  

2-100Hz 

260µs-150µs  Continue

   

Rate & width

   

over 6 seconds

   P12 

Modulation  

80< - >7Hz 

260µs 

Continue

   

Rate & width

   

over 6 seconds

  Program     

The pre-set parameters of the EMS 13-24 

 

 

Details

 

programs

 

are as given below:

 

   

Mode 

SYN/   Rate  Width   Ramp  ON Time Off Time  Timer

   

 

ALT   (Hz)   (µs)    (Sec)  (Sec) 

(Sec)  

(Min)

  P13 

ACL repair/joint 

SYN   35 

300 

24 

20

   

protection back muscle

 

P14 

Spasm small  

SYN   80 

300 

10 

20

   

muscle 

 

P15 

Spasm 

SYN  80 

250 

20 

   

Postoperative

 

P16 

Arthroscopy  

SYN  25 

200 

30 

15

 

P17 

Disuse atrophy  

SYN  35 

300 

15 

30

 

P18 

Shoulder 

SYN  50 

300 

15 

50 

15 

   

Subluxation

 

P19 

Range of motion 

SYN  40 

250 

21 

30 

   

muscle re-education of hips

 

P20 

Muscle training 

SYN  50 

250 

10 

10 

20

 

P21 

Muscle training 

SYN  50 

250 

14 

14 

20

 

P22 

Muscle training 

SYN  35 

400 

10 

10 

20

 

P23 

Muscle training 

ALT  50 

250 

10 

10 

20

  P24 

Muscle training 

ALT  50 

250 

10 

14 

20

  

23   Timer 

Adjustable, from 5 to 60 minutes minutes

           

and continue(C), 5 minutes each step

  24   Patient  

This unit can store 60 sets of operation 

        Compliance Meter  records. Total recorded time is 999 hours. 
  25   Low Battery  

A low battery indicator will show up when   

        Indicator 

the battery is low.  

  26   Operating 

Temperature:0°~40°C

        Condition 

Relative Humidity: 30%~75%

         

Atmosphere Pressure : 700Hpa~1060Hpa 

  27   Remark 

There may be up to a +/-10% tolerance of  

         

all parameters and +/-20% tolerance of  

         

output amplitude & voltage.

Summary of Contents for EM-6300A

Page 1: ...FOR THE INSTRUCTION MANUAL INSTRUCTION MANUAL INSTRUCTION MANUAL Premier Combo Plus DIGITAL TENS EMS Distributed by V1 0 0434 Read Before Use Model No EM 6300A ...

Page 2: ...ctrode Lead Wires 20 15 Lead Wire Maintenance 21 16 Electrode Options 21 17 Electrode Placement 21 18 Tips for Skin Care 22 19 Application of Reusable self adhesive electrodes 22 20 Adjusting the Controls 24 21 Battery Information 32 22 Maintenance Transportation and Storage 34 23 Safety Technical Controls 34 24 Malfunctions 35 25 Conformity to Safety Standards 35 26 Warranty 36 Manufacturer 36 Re...

Page 3: ...reases the muscle contracts as in physical exercise Then when the pulse ceases the muscle relaxes and the cycle starts over again Stimu lation Contraction and Relaxation Powered muscle stimulators Chapter 1 GENERAL DESCRIPTION The Premier Combo Plus TENS EMS is a fully digital battery oper ated pulse generator that sends electrical impulses to the nerves and underlying muscle groups This unit is a...

Page 4: ...mmended for use by the manufacturer 10 Portable powered muscle stimulators should not be used while driving operating machinery or during any activity in which involuntary muscle contractions may put the user at undue risk of injury should only be used under medical supervision for adjunctive thera py for the treatment of medical diseases and conditions IMPORTANT SAFETY INFORMATION Read instructio...

Page 5: ...ay be strong enough to close the airway or cause difficulty in breathing 4 Stimulation should not be applied transthoracically in that the introduction of electrical current into the heart may cause cardiac arrhythmias 5 Stimulation should not be applied transcerebrally 6 Stimulation should not be applied over swollen infected or inflamed areas or skin eruptions e g phlebitis thrombophlebitis vari...

Page 6: ...table 1 8 seconds 1 Sec step The On time will increase and decrease in the setting value 13 Mode Six TENS Modes B Burst N Normal M Modulation Rate Width S1 Modulation Width S2 Modulation Width and P Three EMS Modes C Constant S Synchronous A Alternate 14 Burst Mode Burst rate Adjustable 0 5 5Hz Pulse width adjustable 50 300µs Frequency fixed 100 Hz 15 Normal Mode The pulse rate and pulse width are...

Page 7: ...s 1 second In this mode pulse rate 2 150Hz and pulse width 50 300µs are fully adjustable 17 S1 Mode Pulse width is automatically varied in a cyclic pattern over a nominal 10 second period Pulse width decreases over a period of 4 seconds from the initial setting to a value 40 less The narrower pulse width continues for 1 second It then increases over a period of 4 seconds to its initial setting The...

Page 8: ...all SYN 80 300 3 10 5 20 muscle P15 Spasm SYN 80 250 2 8 4 20 Postoperative P16 Arthroscopy SYN 25 200 2 6 30 15 P17 Disuse atrophy SYN 35 300 2 5 15 30 P18 Shoulder SYN 50 300 5 15 50 15 Subluxation P19 Range of motion SYN 40 250 3 6 21 30 muscle re education of hips P20 Muscle training SYN 50 250 2 10 10 20 P21 Muscle training SYN 50 250 2 14 14 20 P22 Muscle training SYN 35 400 2 10 10 20 P23 M...

Page 9: ... replace the other parts of a device PARTS 1 Lead Wires 2 Electrodes 3 Belt Clip 4 Lead Connector 5 Main PCB 6 Press Buttons 7 LCD Cover 8 9V Battery 9 Device Case Chapter 10 ACCESSORIES Each Premier Combo Plus Digital TENS EMS comes complete with standard accessories and the standard labels as given below I Accessories REF NO DESCRIPTION Q TY 1 KF4040 40x40mm Adhesive Electrodes 4 pieces 2 KE 24 ...

Page 10: ...007 47 EC Notify body Det Norske Veritas DNV 15 Power Chapter 12 OPERATING INSTRUCTIONS 1 Insert the 9V battery into the device s battery compartment Make sure that the plastic seal on the 9V battery is removed Line up the positive and negative terminals on the battery with their corresponding terminals in the device Make sure that the unit is turned off 2 Insert the lead wires into the lead wire ...

Page 11: ...rve accommodation by continuously cycling the treatment intensity Advise the patient to increase the intensity very slowly when using modulation mode INTENSITY Each patient responds differently to different levels of intensity due to varying degrees of tissue resistance enervation skin thickness etc Intensity instructions are therefore limited to the following settings Perception The intensity is ...

Page 12: ...he jacks see drawing one or two sets of wires may be used After connecting the wires to the stimulator attach each wire to an electrode Use care when you plug and unplug the wires Jerking the wire instead of holding the insulated connector body may cause wire breakage CAUTION Do not insert the plug of the patient lead wire into any AC power supply socket Chapter 15 LEAD WIRE MAINTENANCE Clean the ...

Page 13: ...trode placement area when not wearing electrodes 8 Never apply electrodes over irritated or broken skin Chapter 19 APPLICATION OF RE USABLE SELF ADHESIVE ELECTRODES Application 1 Clean and dry the skin at the prescribed area thoroughly with soap and water prior to application of electrodes 2 Insert the lead wire into the pin connector on the pre wired electrodes 3 Remove the electrodes from the pr...

Page 14: ...ct with the skin Pause 3 Mode Control There are 5 TENS modes B N M S1 S2 and 3 EMS modes C S A available The mode is selected by pressing the Mode control When a TENS mode is selected the LCD shows TENS When EMS mode is selected the LCD shows EMS 4 Set Control By pressing the Set control you select the setting you intend to adjust The value is set by pressing the Increment or Decrement controls wh...

Page 15: ...be stored immediately after selected There are 12 preset programs of your option on P mode Press or to select a program desired after P mode is selected These programs are not adjustable c Set Pulse Width Pulse Width is adjustable from 50 µs to 300 µs Press SET control to enter this menu then press or to adjust the setting If no instructions regarding the pulse width are given in therapy set the c...

Page 16: ...stal display After a mode is selected always press Set to enter next setting and press or to adjust its value The settings will be stored immediately after selected After a mode is selected always press Set to enter next setting and press or to adjust its value The settings will be stored immediately after selected There are 12 preset programs of your option on P mode Press or to select a program ...

Page 17: ...Hz range h Set Timer The treatment time is adjustable between 5 60 minutes and Continue C Press SET control to enter this menu then press Increment or Decrement to adjust the setting The liquid crystal will show the balance treatment time after the stimulation is started Output will be terminated when time is up Turn off the unit when the output is off i Adjust Intensity There are 99 steps within ...

Page 18: ...g periods of storage batteries should be charged prior to use BATTERY CHARGING 1 Plug the charger into any working 110 or 220 240v mains electrical outlet The use of any attachment not supplied with the charger may result in the risk of fire electric shock or injury to persons 2 Follow the battery manufacturer s instructions for charging time 3 After the battery manufacturer s recommended charging...

Page 19: ...Combo Plus DIGITAL TENS EMS 1 Check the device for external damage deformation of the housing damaged or defective output sockets 2 Check the device for defective operating elements legibility of inscriptions and labels make sure the inscriptions and labels are not distorted 3 Check the usability of accessories patient cable undamaged electrodes undamaged Battery is not corroded Please consult you...

Page 20: ...t cause any interference with other devices in the proximity It is recommended that you do not place the device on top of or close to other electronic devices Should you notice any inter ference with other electrical devices move the device or connect it to a different socket Radio equipment may affect the operation of this device Guidance and manufacturer s declaration electromagnetic emissions T...

Page 21: ... the electromagnetic environment specified below The customer or the user of the EM 6300A should assure that it is used in such an environment IMMUNITY test Conducted RF IEC 61000 4 6 Radiated RF IEC 61000 4 3 IEC 60601 test level 3 Vrms 150 kHz to 80 MHz 3 V m 80 MHz to 2 5 GHz Compliance level 3 Vrms 3 V m Electromagnetic environment guidance Portable and mobile RF communications equipment shoul...

Page 22: ...paration distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electro magnetic propagation is affected by absorption and reflection from structures objects and people Recommended separation distances between portable and mobile RF communications equipment and the EM 6300A The EM 6300A is intended for use in an electromagnetic environment in which ...

Page 23: ...V pk pk Pulse Width 300µs 2 N MODE Normal Load 500 ohm Pulse Rate 150Hz Pulse Width 300µs Appendix II Waveform of EM 6300A Digital TENS EMS TENS 1 B Mode Burst Load 500 ohm Pulse Rate 150Hz Pulse Width 300µs Scope A VERT 10 0V DIV HORIZ 2mS OUTPUT 57 1875Vpk pk Pulse Rate 100Hz Scope B VERT 10 0V DIV HORIZ 50mS Pulse Rate 5 000Hz ...

Page 24: ... 150µs 3 M MODE 50 Pulse Width Rate Modulation Load 500 ohm Pulse Rate 150Hz Pulse Width 300µs Scope C Modulation 50 Scope A VERT 20 0V DIV HORIZ 200 µs OUTPUT 50 0V pk pk Pulse width 184 µs Scope B VERT 20 0V DIV HORIZ 200 µs OUTPUT 56 9V pk pk Pulse width 302 µs 4 S1 MODE 40 Pulse Width Modulation Load 500 ohm Pulse Rate 150Hz Pulse Width 300 µs ...

Page 25: ...6 9Vpk pk Pulse width 302 µs 5 S2 MODE 70 Pulse Width Modulation Load 500 ohm Pulse Rate 150Hz Pulse Width 300µs Scope A VERT 10 0V DIV HORIZ 2mS OUTPUT 59 1825V pk pk Pulse Rate 148 8Hz Scope B VERT 10 0V DIV HORIZ 100µs OUTPUT 59 1825V pk pk PulseWidth 300µs EMS 1 C MODE Load 500 ohm Pulse Rate 150Hz Pulse Width 300 µs ...

Page 26: ...00 ohm Pulse Rate 150Hz Pulse Width 300µs Contratction Time 12 Sec Relation Time 12 Sec Ramp Time 6 Sec 3 A MODE Alternate Load 500 ohm Pulse Rate 150Hz Pulse Width 300µs Contratction Time 12 Sec Relation Time 12 Sec Ramp Time 6 Sec ...

Reviews: