27
26
being changed but the output may be stopped by
pressing the "On/Off/Pause".
10. Steps to Set a TENS Program
The settings can be adjusted as follows
a. Turn on the Power
After the electrodes are placed firmly on skin and
the lead wires are plugged in the socket of device,
turn the unit on by pressing the On/Off/Pause
button. The settings will be displayed on the LCD
screen.
b. Select a Mode or a Preset Programme
Select a mode by pressing the "Mode" control.
The mode you selected will show up on the top of liquid
crystal display. There are 5 modes of your option including -
B(Burst), M(Normal), M(Modulation), S1, S2 and P. When a
TENS mode is selected, it shows "TENS" on the liquid crystal
display.
After a mode is selected, always press "Set" to enter next
setting, and press "
" or "
" to adjust its value.
The settings will be stored immediately after selected.
There are 12 preset programs of your option on P mode.
Press "
" or "
" to select a program desired
after P mode is selected. These programs are not adjustable.
c. Set Pulse Width
Pulse Width is adjustable from 50 µs to 300 µs. Press “SET”
control to enter this menu, then press “
” or “
“
to adjust the setting. If no instructions regarding the pulse
width are given in therapy, set the control to the suggested
70-120 µs setting.
d. Set Pulse Rate
Pulse rate is adjustable from 2Hz to 150 Hz . Press "SET"
control to enter this menu, then press "
" or "
"
to adjust the setting. Unless otherwise instructed, set the
pulse rate l to the 70-120 Hz range.
e. Set Timer
The treatment time is adjustable between 5 - 60 minutes and
Continue(C). Press "SET" control to enter this menu, then
press "
" or "
" to adjust the setting. The liquid
crystal will show the balance treatment time after the
stimulation is started. Output will be terminated when time
is up. Turn off the unit when the output is off.
Continuous End of Treatment