11
10
time. The OFF Time should be equal to or
greater than the ON Time.
ON TIME
≥
Ramp up + Ramp down
OFF TIME
≥
ON TIME
22 Mode P
The pre-set parameters of the 24
programs are as given below:
Program
The pre-set parameters of the TENS 1-12
Details
programs
are as given below:
Program Mode Pulse Rate Pulse Width Timer
P1 Constant
80Hz 180µs
Continue
P2 Burst 2Hz
180µs Continue
P3
P.W.
80Hz
70µs -180µs Continue
Modulation
P4
Mixed
15Hz in 3 Sec / 180µs
Continue
Frequency
2Hz in 3 Sec
P5
Mixed
80Hz in 3 Sec / 180µs
Continue
Frequency 2Hz in 3 Sec
P6 Constant
10Hz
180µs
Continue
P7 Constant 80Hz
60µs Continue
P8 Constant
10Hz
200µs
30 Minutes
P9 Burst 50Hz
250µs 30 Minutes
P10
on Time Pause Time Frequency Pulse Width Sub-
Active
Pulse
(sec) (sec) (Hz)
(µs)
Pulse Channels
Polarity
2
0
5
120
No
CH1 & CH2 Positive
1.5
0
8
120
No
CH1 & CH2 Positive
1.4
0
10
120
No
CH1 & CH2 Positive
1
0
15
120
No
CH1 & CH2 Positive
0.8
0
20
120
No
CH1 & CH2 Positive
0.7
0
30
120
No
CH1 & CH2 Positive
0.7
0
40
120
No
CH1 & CH2 Positive
0.7
0
50
120
No
CH1 & CH2 Positive
1
0
75
120
No
CH1 & CH2 Positive
in 0.5 second, then the pulse rate is
decreased by 50% from its original setting
in 0.5 second. Total cycle time is 1 second.
In this mode, pulse rate(2-150Hz) and
pulse width(50-300µs) are fully adjustable.
17 S1 Mode
Pulse width is automatically varied in a
cyclic pattern over a nominal 10 second
period. Pulse width decreases over a
period of 4 seconds from the initial setting
to a value 40% less.
The narrower pulse width continues for 1
second. It then increases over a period of
4 seconds to its initial setting. The cycle is
then repeated. Pulse rate and pulse width
are fully adjustable.
18 S2 Mode
Pulse width is automatically varied in a
cyclic pattern over a nominal 10 second
period. Pulse width decreases over a
period of 4 seconds from the initial setting
to a value 70% less.
The narrower pulse width continues for 1
second. It then increases over a period of
4 seconds to its initial setting. The cycle is
then repeated. Pulse rate and pulse width
are fully adjustable.
19 Constant
The pulse rate and pulse width are
Mode(C)
adjustable. It generates continuous
stimulation is delivered .
20 Synchronous
Output from both channels occurs
Mode(S)
synchronously. The "ON" time includes
"Ramp Up" and "Ramp Down" time.
Therefore, the setting of ON Time should
be no less than two times of the "Ramp"
time in this mode.
21 Alternate
The stimulation of the CH2 will occur after
Mode(A)
the 1st contraction of CH1 is completed. In
this mode, the setting of ON Time should
be no less than two times of the "Ramp"