31
30
e. Set Off Time
The Off Time controls the length of relaxation. By pressing
the "SET" control, the relaxation time can be adjusted. Both
channels' stimulation is cycled on and off by the contraction
and relaxation settings. The range is adjustable from 2
seconds to 90 seconds.
In Alternate mode, the OFF Time should be equal or more than
the ON Time. (OFF TIME
≥
ON TIME)
f. Set Pulse Width
Pulse Width is adjustable from 50 µs to 300 µs. Press "SET"
control to enter this menu, then press "
" or "
"
to adjust the setting. If no instructions regarding the pulse
width are given in therapy, set the control to the suggested
70-120 µs setting
g. Set Pulse Rate
Pulse rate is adjustable from 2Hz to 150 Hz . Press "SET"
control to enter this menu, then press "
"or "
"
to adjust the setting.
Unless otherwise instructed, set the pulse rate to the 70-
120 Hz range.
h. Set Timer
The treatment time is adjustable between 5 - 60 minutes and
Continue(C). Press "SET" control to enter this menu, then
press "Increment" or "Decrement" to adjust the setting. The
liquid crystal will show the balance treatment time after the
stimulation is started. Output will be terminated when time is
up. Turn off the unit when the output is off.
i. Adjust Intensity
There are 99 steps within the intensity range. Set the desired
level by pressing the " " or " " controls. Press the
"Lock" button to prevent accidental changes.
12. Compliance Meter
The individual treatment time and total treatment time can be
checked and deleted by the following steps. Sixty sets of
treatment records can be stored. Total recorded time is 999
hours.
Check & Delete Treatment Record
Press "Mode" control and turn on the power simultaneously.
The LCD will show the individual operation time. Press "Mode"
control to check the accumulated treatment time. The record
can be deleted by pressing the "SET" button for two seconds.
Continuous End of Treatment