f. Adjust Intensity
There are 99 steps within the intensity range. Set the desired
level by pressing the " " or " " controls. Press the
"Lock" button to prevent accidental changes.
11. Steps to Set a EMS Program
The settings can be adjusted according to the following steps.
a. Turn on the Power
After the electrodes are placed firmly on skin and the lead
wires are plugged in the socket of device, turn the unit on by
pressing the On/Off/Pause button. The settings will show
up on LCD for your further adjustment.
b. Select a Mode or a Preset Programme
Select a mode by pressing the "Mode" control. The mode you
selected will show up on the top of liquid crystal display.
There are 3modes of your option including -
C(Constant), S(Synchronous), A(Alternate). When an EMS
mode is selected, it shows "EMS" on the liquid crystal display.
After a mode is selected, always press "Set" to enter next
setting, and press "
" or "
" to adjust its value.
The settings will be stored immediately after selected.
After a mode is selected, always press “Set” to enter next
setting, and press “
“ or “
“ to adjust its value.
The settings will be stored immediately
after selected.
There are 12 preset programs of your option on P mode.
Press “
“ or “
“ to select a program desired
after P mode is selected. These programs are not adjustable.
c. Set Ramp Time
The ramp time controls the time taken to reach maximum and
the time taken to fall to zero I order to make the contraction
more comfortable. The ramp time is adjustable between 1 - 8
seconds.
d. Set On Time
The On Time controls the length of stimulation. By pressing
the "Set" control, the contraction time can be adjusted. Both
channels' stimulation is cycled on and off by the contraction
and relaxation settings. The range is adjustable from 2
seconds to 90 seconds.
The total "ON" time must be at least twice the "Ramp" time