User’s Guide
Note:
If
AUTO RETURN
is turned
on
,
CYCLING
is automatically
turned
off
and the
KEYS
parameter is automatically set to
MAIN-
TAINED
mode. See following pages for details on the
CYCLING
and
KEYS
parameters.
5.3 Cycling (CYCLES)
This setting allows the user to configure the number of up and down
cycles through which the crosshead will sequence. One cycle con-
sists of the crosshead moving to a limit switch, force set point, or
travel distance, whichever occurs first, at the specified speed, stop-
ping for the specified amount of dwell time, and returning to the
other limit at the specified speed. If the independent up and down
speed feature is not enabled, the speed will be the same in both
directions.
Default setting: 00000 (off)
Available settings: 00000 – 99999
AUTO RETURN:
o f f
ESC
ENTR
+
_
+ or –
Cycles through the available settings
ENTR
Returns to the Test Parameter Setup menu
ESC
Exits the parameter without saving changes
C Y C L E S :
0 0 0 0 0
ES C
ENTR
+
_
12