Abbreviated Operator’s Guide 

CS300

 |

n

 

 With sustained unpredictable rhythms and a valid ECG Trigger, R-wave deflation 
is automatically selected. The message 

“Auto R-Wave Deflate”

 will be displayed. 

The deflation indicator will automatically be repositioned to the far right, “R-Trac” will 
be displayed in the IAB deflation indicator. The system will return to predictive timing 
once the rhythm becomes predictable.

n

  Loss of trigger causes an alarm and pumping will stop. 

Manual Operation Mode

n

   Trigger source and timing of IAB inflation and deflation is determined by the operator.

n

   Changing the trigger source will cause the pump to go to Standby. Pumping 

will resume when the 

START

 key is pressed.

n

  Leads I, II, III, AVR, AVL, AVF, V or External ECG source can be selected.

n

  Timing must be readjusted by the operator if heart rate or rhythm changes.

n

  Loss of trigger causes an alarm and pumping will stop.

n

  Manual timing is typically used for 

pediatric

 IABP patients.

WARNING:

 The CS300 must be in Semi Auto Operational mode whenever no aortic 

pulse is present, and IAB assist is desired. For example, whenever circulatory bypass 
or a laminar flow, left ventricular assist device is in use.

Trigger Source

n

  Trigger is the signal the CS300 uses to identify the beginning of the cardiac cycle.

n

   The Trigger Source keys are only active in the 

SEMI AUTO and Manual Operation 

Modes

.

n

   Regardless of the Operation Mode, the active trigger source is indicated by a flashing 

diamond in the upper right corner of the screen.

ECG

n

  The R wave of the ECG is the trigger event.

n

  Pacer spikes are rejected.

n

  ESIS (electrosurgical interference suppression) is automatic in this mode.

n

   When using SEMI AUTO or Manual Operation modes, the ECG gain can be adjusted 

from 0.15 to 3.00, relative to a normal gain of 1.0. The amount of gain will be 
displayed next to ECG GAIN on the monitor screen. Go to Pump Options – ECG Gain 
to adjust.

Pressure

n

  The systolic upstroke of the arterial waveform is the trigger event.

n

 

 The CS300 will automatically adapt the pressure trigger threshold to the systolic 
pulse height of the arterial pressure waveform.

n

   The pressure trigger threshold is displayed just below the trigger source on the 

monitor screen and the arterial waveform is marked with a horizontal tick to indicate 
the trigger point.

VAULT COPY

Summary of Contents for Datascope CS300

Page 1: ...DATASCOPE IS NOW MAQUET CARDIOVASCULAR THE CS300 ABBREVIATED OPERATOR S GUIDE VAULT COPY...

Page 2: ...CS300 Bezel and Keypad The CS300 provides comprehensive HELP SCREENS with easy to follow step by step instructions on set up as well as alarm and advisory conditions VAULT COPY...

Page 3: ...nsor IAB Calibration On initial insertion pressing the START key automatically purges and fills the IAB and calibrates the Sensor IAB if connected While assisting press and hold the Zero Pressure key...

Page 4: ...ATING MODES AUTO Operation Mode All aspects of pump Operation are automated n Most appropriate trigger source is automatically selected n Inflation and deflation timing is automatically set n Lead I I...

Page 5: ...d IAB assist is desired For example whenever circulatory bypass or a laminar flow left ventricular assist device is in use Trigger Source n Trigger is the signal the CS300 uses to identify the beginni...

Page 6: ...he A V interval is between 80 224 msec and the pacing rate is less than 125 beats per minute Pacer A n This trigger mode is only available in the SEMI AUTO and Manual Operation modes n The R wave of t...

Page 7: ...eration mode n Use the IAB Deflation arrows to adjust the end of the highlighted segment prior to ventricular ejection After initiating assist the IAB INFLATION only operational in the Semi Auto and M...

Page 8: ...is sensed In SEMI AUTO Operation mode if the ectopic beat is of small amplitude reliable triggering can be maximized if an ECG Lead is selected which minimizes the amplitude difference between the nor...

Page 9: ...energy coupled from the ESU to the ECG input of the CS300 generally resulting in stable ECG triggering When the CS300 is in AUTO Operation mode it will automatically select Pressure Trigger when ESU i...

Page 10: ...or ALARM messages are displayed in the ALARM MESSAGES section of the monitor display HIGH PRIORITY ALARMS High Priority Alarms indicate situations that require immediate Operator response Pumping is s...

Page 11: ...o improve ECG signal by ensuring electrode contact and optimal placement n If signal problem persists switch Operation mode to SEMI AUTO Verify trigger source n Use IAB Inflation and IAB Deflation con...

Page 12: ...rigger is required and manual threshold is selected reduce the pressure trigger threshold by using the PRESSURE THRESHOLD sub menu from the PUMP OPTIONS menu n If appropriate select a different trigge...

Page 13: ...M The IAB fill mode has been changed from Manual to AUTO without an autofill Operator Intervention n Inspect the IAB and catheter extender tubing for any evidence of a leak including the safety disk c...

Page 14: ...Check integrity of all cable lead connections n Attempt to improve Arterial Pressure by checking integrity of all cable connections If conventional transducer in use verify the transducer was not lef...

Page 15: ...no fill had been performed Verify Proper Timing SA M Operator has changed the trigger source Manual Fill IAB SA M Notifies the operator when to manually fill the IAB catheter with shuttle gas Display...

Page 16: ...ed the Semi Auto Operation mode Maintenance Required Code ___ System maintenance may be required Low Helium Helium supply is below a preset reserve as determined by tank pressure Prolonged Time in Sta...

Page 17: ...elect an alternative item use the UP and DOWN arrow keys 24hr Emergency Helpline 1 800 777 4222 Sweep Speed Balloon Waveform Brightness ECG Inflation Markers Alarm Volume Standby Advisory Tone Beep Vo...

Page 18: ...ous printout hold the PRINT STRIP key for two 2 seconds The associated LED will blink as printing continues To stop the continuous printout press the PRINT STRIP key again Doppler Information The dopp...

Page 19: ...p the IAB Catheter Extender Input when the message PLUG DISK OUTLET appears on the display If the message UNPLUG DISK OUTLET appears remove any connections that may exist to the IAB Catheter Extender...

Page 20: ...The condensate removal module will not operate to conserve remaining power Battery Charging To charge the internal battery 1 Leave the system power cord plugged in and set the MAINS On Off to ON 2 Ch...

Page 21: ...t and that a noticeable pressure drop has not occurred NOTE Prior to using the system in any other position except vertical consult with Technical Service NOTE For information on outfitting an aircraf...

Page 22: ...off the cart To remove the console by detaching the battery pack With the console on the cart lift the battery release levers to the unlock position fig 2 Lift the console straight up and off the cart...

Page 23: ...e by pressing the button on the rear of the monitor fig 4 Attach the monitor to the top of the pump console Make sure the monitor is securely attached before moving the system Extend the console handl...

Page 24: ...o technical development Caution U S Federal Law restricts this device to sale by or on the order of a physician Refer to Instructions for Use for current indications warnings contraindications and pre...

Reviews: