MEDIUM PRIORITy ALARMS
Medium Priority Alarms indicate situations in which a prompt Operator response is
required. This class of alarm does not suspend pumping, but may indicate a need for
corrective action. A yellow flashing alarm icon is displayed as seen above. The audio tone
for a Medium Priority Alarm is three notes, a pause, and then this cycle repeats.
n
Status/Prompts apply to all Operation modes unless otherwise indicated:
A
= Auto /
SA
= Semi Auto /
M
= Manual Operation mode.
ALARM MESSAGE
PROBABLE CAUSE
Poor Signal Quality
[A]
Both ECG and Arterial Pressure signal quality
are poor while in ECG trigger.
No Pressure Source Available
[A]
Neither a direct nor external blood pressure
source was detected.
IAB Optical Sensor Failure
[ALL]
There has been a failure of the Optical Sensor
in the IAB.
Low Battery
[ALL]
Battery reserve falls below 30 minutes of
operating time.
OPERATOR INTERVENTION
for Poor Signal Quality
n
Attempt to improve ECG signal by ensuring electrode contact and optimal placement.
Check integrity of all cable/lead connections.
n
Attempt to improve Arterial Pressure by checking integrity of all cable connections. If
conventional transducer in use, verify the transducer was not
left vented and/or flush the lumen per standard hospital procedure.
for No Pressure Source Available
n
If using a Sensor IAB, ensure that the sensor cable has been connected. Otherwise
provide an A.P. source by connecting a standard transducer or an interface cable
from the A.P. output of an external monitor to the pump.
for IAB Optical Sensor failure
n
Ensure that the sensor cable has been connected.
n
Provide an alternate A.P. source by connecting a standard transducer or an interface
cable from the A.P. output of an external monitor to the pump.
for Low Battery
n
Verify the Mains Power Switch, located above the AC power cord connector, is ON.
n
Connect the system to an AC power outlet.
LOW PRIORITy ALARM
Low Priority Alarms are indications that Operator awareness is required. A yellow
flashing alarm icon is displayed as seen above. The audio tone for a low priority alarm
is two notes, a pause, and then this cycle repeats.
ALARM MESSAGE
[AUTO OPERATION MODE]
PROBABLE CAUSE
Unable to Update Timing
Waveform quality is poor.
VAULT COPY