10
|
CS300
™
|
Abbreviated Operator’s Guide
|
OPERATOR INTERVENTION
for No Trigger
n
Reattach or reposition the electrode and check the integrity of all cable/lead
connections.
n
Try ECG lead selections via the ECG/AP Sources menu, or select EXT. ECG,
if appropriate.
n
Adjust the ECG Gain by accessing the
PUMP OPTIONS
menu.
n
Select a different trigger source using the
TRIGGER SOURCE
arrow keys. Resume
pumping by pressing the
START
key.
for No Pressure Trigger
n
Verify that the desired Pressure Source, DIRECT/EXTERNAL, is selected. Use
the ECG/AP Sources menu to select the Pressure Source.
n
Verify that the transducer was not left vented; check all transducer cable connections.
n
If pressure trigger is required and manual threshold is selected, reduce the pressure
trigger threshold by using the
PRESSURE THRESHOLD
sub menu from the
PUMP
OPTIONS
menu.
n
If appropriate, select a different trigger source using the TRIGGER SOURCE arrow
keys. Resume pumping by pressing the
START
key.
for Check Pacer timing:
n
Select ECG trigger source, using the TRIGGER SOURCE arrow keys, when demand
pacing a patient.
n
Reduce the Pacer rate.
n
Adjust the A-V time interval to between 80 and 224 msec.
for Trigger Interference:
n
Pumping automatically resumes when interference clears.
n
If the condition persists, then select pressure trigger using the
TRIGGER SOURCE
arrow keys.
n
Verify timing and press the
START
key to resume pumping.
for ECG Detected
n
Select ECG as the trigger source using the TRIGGER SOURCE arrow keys.
n
Verify proper timing.
n
Resume pumping by pressing the
START
key.