CS300
™
|
Abbreviated Operator’s Guide
|
IAB Sensor Low Level Pressure Output – to Bedside Monitor
This output connector is used to provide an electrical signal from the IAB Sensor
module to the physiologic pressure transducer input on the patient’s bedside monitor
when a Sensor IAB is connected and calibrated. Its low level output emulates that of a
typical Arterial Pressure Transducer, allowing direct connection to the patient inputs of
the bedside monitor when a Sensor IAB is connected and calibrated.
IAB Sensor Output – Vent
This button facilitates the zeroing process required for proper calibration of the
physiologic pressure channel of an attached bedside monitor. When pressed, the IAB
Sensor Output signal will simulate that of an atmospheric vented transducer. Venting
will be simulated for 15 seconds to give users adequate time to zero the pressure
channel on the bedside monitor.
The pressure signal will then be restored automatically to permit monitoring
through the bedside and centralized monitoring systems.
OPERATING MODES
AUTO Operation Mode
All aspects of pump Operation are automated
n
Most appropriate trigger source is automatically selected.
n
Inflation and deflation timing is automatically set.
n
Lead I, II, III, or External ECG source is automatically selected.
n
System searches for the next best trigger source in the event of a loss of
trigger source.
n
Timing adjusts automatically if trigger source changes, heart rate changes,
rhythm changes.
n
With sustained unpredictable rhythms, R-wave deflation is automatically selected.
The message
“Auto R-Wave Deflate”
will be displayed. The deflation indicator
will automatically be repositioned to the far right, “R-Trac” will be displayed in the
IAB deflation indicator. The system will return to predictive timing once the rhythm
becomes predictable.
SEMI AUTO Operation Mode
n
Operator selects the most appropriate trigger source.
n
Operator establishes initial timing and thereafter, software algorithms automatically
track changes in patient heart rate or rhythm and adjusts timing accordingly.
n
Changing the trigger source will cause the pump to go to Standby. Pumping will
resume when the
START
key is pressed.
n
Lead I, II, III, AVR, AVL, AVF, V or External ECG source can be selected.