14
|
CS300
™
|
Abbreviated Operator’s Guide
|
System Test Ok
All systems passed power-up diagnostics.
System Trainer
Displayed when the System Trainer is being
used as ECG and Pressure source.
Auto R-Wave Deflate
[A /SA]
The patient’s rhythm is too random to predict
the occurrence of the next beat. In response,
the system has automatically selected the
R-Wave Deflation Mode.
R-Wave Deflate
[A /SA]
Operator has enabled R-Wave deflation by
moving the deflate key to the extreme right
position.
Irregular Pressure Trigger
[SA]
While Pressure is the Trigger source, erratic
trigger is due to:
1. Arrhythmic patient rhythm.
2. Late deflation; inhibiting pressure pulse
detection.
Manual Timing selected – See Help
[M]
The operator selected the Manual Operation
mode.
Verify Proper Timing
[SA]
The operator selected the Semi Auto
Operation mode.
Maintenance Required Code # ___
System maintenance may be required.
Low Helium
Helium supply is below a preset reserve as
determined by tank pressure.
Prolonged Time in Standby
Pump has been in STANDBY for at least 10
minutes.
A.P. Optical Sensing Module Failure
There has been a failure of the A.P. Optical
Sensing Module in the pump console.
Unable to Calibrate IAB Optical Sensor
A calibration is not completed for one or
more of the following reasons:
– the patient’s arterial pressure is too low
to reliably establish or update IAB Optical
Sensor Calibration
– there is a restriction in the catheter
– the IAB Fill Mode is set to Manual
IAB Optical Sensor Calibration Expired
A calibration update has been intentionally
postponed for one or more of the following
reasons:
– the patient’s Mean Arterial Pressure may be
too low
– less than 15 minutes have elapsed since
last calibration
– pump is either in STANDBY or the IAB FILL
mode is set to MANUAL