ABBREVIATED OPERATOR’S GUIDE
VAULT COPY
Page 1: ...DATASCOPE IS NOW MAQUET CARDIOVASCULAR THE CS300 ABBREVIATED OPERATOR S GUIDE VAULT COPY...
Page 2: ...CS300 Bezel and Keypad The CS300 provides comprehensive HELP SCREENS with easy to follow step by step instructions on set up as well as alarm and advisory conditions VAULT COPY...
Page 3: ...nsor IAB Calibration On initial insertion pressing the START key automatically purges and fills the IAB and calibrates the Sensor IAB if connected While assisting press and hold the Zero Pressure key...
Page 4: ...ATING MODES AUTO Operation Mode All aspects of pump Operation are automated n Most appropriate trigger source is automatically selected n Inflation and deflation timing is automatically set n Lead I I...
Page 5: ...d IAB assist is desired For example whenever circulatory bypass or a laminar flow left ventricular assist device is in use Trigger Source n Trigger is the signal the CS300 uses to identify the beginni...
Page 6: ...he A V interval is between 80 224 msec and the pacing rate is less than 125 beats per minute Pacer A n This trigger mode is only available in the SEMI AUTO and Manual Operation modes n The R wave of t...
Page 7: ...eration mode n Use the IAB Deflation arrows to adjust the end of the highlighted segment prior to ventricular ejection After initiating assist the IAB INFLATION only operational in the Semi Auto and M...
Page 8: ...is sensed In SEMI AUTO Operation mode if the ectopic beat is of small amplitude reliable triggering can be maximized if an ECG Lead is selected which minimizes the amplitude difference between the nor...
Page 9: ...energy coupled from the ESU to the ECG input of the CS300 generally resulting in stable ECG triggering When the CS300 is in AUTO Operation mode it will automatically select Pressure Trigger when ESU i...
Page 10: ...or ALARM messages are displayed in the ALARM MESSAGES section of the monitor display HIGH PRIORITY ALARMS High Priority Alarms indicate situations that require immediate Operator response Pumping is s...
Page 11: ...o improve ECG signal by ensuring electrode contact and optimal placement n If signal problem persists switch Operation mode to SEMI AUTO Verify trigger source n Use IAB Inflation and IAB Deflation con...
Page 12: ...rigger is required and manual threshold is selected reduce the pressure trigger threshold by using the PRESSURE THRESHOLD sub menu from the PUMP OPTIONS menu n If appropriate select a different trigge...
Page 13: ...M The IAB fill mode has been changed from Manual to AUTO without an autofill Operator Intervention n Inspect the IAB and catheter extender tubing for any evidence of a leak including the safety disk c...
Page 14: ...Check integrity of all cable lead connections n Attempt to improve Arterial Pressure by checking integrity of all cable connections If conventional transducer in use verify the transducer was not lef...
Page 15: ...no fill had been performed Verify Proper Timing SA M Operator has changed the trigger source Manual Fill IAB SA M Notifies the operator when to manually fill the IAB catheter with shuttle gas Display...
Page 16: ...ed the Semi Auto Operation mode Maintenance Required Code ___ System maintenance may be required Low Helium Helium supply is below a preset reserve as determined by tank pressure Prolonged Time in Sta...
Page 17: ...elect an alternative item use the UP and DOWN arrow keys 24hr Emergency Helpline 1 800 777 4222 Sweep Speed Balloon Waveform Brightness ECG Inflation Markers Alarm Volume Standby Advisory Tone Beep Vo...
Page 18: ...ous printout hold the PRINT STRIP key for two 2 seconds The associated LED will blink as printing continues To stop the continuous printout press the PRINT STRIP key again Doppler Information The dopp...
Page 19: ...p the IAB Catheter Extender Input when the message PLUG DISK OUTLET appears on the display If the message UNPLUG DISK OUTLET appears remove any connections that may exist to the IAB Catheter Extender...
Page 20: ...The condensate removal module will not operate to conserve remaining power Battery Charging To charge the internal battery 1 Leave the system power cord plugged in and set the MAINS On Off to ON 2 Ch...
Page 21: ...t and that a noticeable pressure drop has not occurred NOTE Prior to using the system in any other position except vertical consult with Technical Service NOTE For information on outfitting an aircraf...
Page 22: ...off the cart To remove the console by detaching the battery pack With the console on the cart lift the battery release levers to the unlock position fig 2 Lift the console straight up and off the cart...
Page 23: ...e by pressing the button on the rear of the monitor fig 4 Attach the monitor to the top of the pump console Make sure the monitor is securely attached before moving the system Extend the console handl...
Page 24: ...o technical development Caution U S Federal Law restricts this device to sale by or on the order of a physician Refer to Instructions for Use for current indications warnings contraindications and pre...