Operation Manual easyScreen
8508935 Rev. 4
14
04/04/2019
In Case of Emergency
In case of emergency, disconnect the device from power supply.
Do not position the device in a way that it is difficult to
operate the disconnection device. The supply mains and the
power socket shall be accessible at all times.
Do not use the device if the mains cable and/or the plug is
damaged.
To transfer data to a PC, establishing a PC-connection via USB
is required. See section 4.4.10 on how to safely establish a
connection with a power supplied PC or laptop (medical
device/non-medical device) or to a battery-driven laptop.
This equipment is intended to be connected to other equipment
thus forming a Medical Electrical System. External equipment
intended for connection to signal input, signal output or other
connectors shall comply with the relevant product standard e.g.
IEC 60950-1 for IT equipment and the IEC 60601-series for
medical electrical equipment. In addition, all such combinations
– Medical Electrical Systems – shall comply with the safety
requirements stated the general standard IEC 60601-1, edition
3, clause 16. Any equipment not complying with the leakage
current requirements in IEC 60601-1 shall be kept outside the
patient environment i.e. at least 1.5 m from the patient support
or shall be supplied via a separation transformer to reduce the
leakage currents. Any person who connects external equipment
to signal input, signal output or other connectors has formed a
Medical Electrical System and is therefore responsible for the
system to comply with the requirements. If in doubt, contact
qualified medical technician or your local representative. If the
device is connected to a PC (IT equipment forming a system)
ensure not to touch the patient while operating the PC. Do not
touch the patient and the printer at the same time.
If the device is connected to a PC (IT equipment forming a
system) assembly and modifications shall be evaluated by
qualified medical technician according to safety regulations
in IEC 60601.
The consequence of not following this warning could be a
too high leakage current to the patient.
Do not touch the contacts on the bottom of the device and the
patient at the same time. When connected to computer
equipment do not simultaneously touch the computer
equipment and the patient at the same time. The consequence
could be a too high leakage current to the patient.
WARNING
WARNING
WARNING