Operation Manual easyScreen
8508935 Rev. 4
59
04/04/2019
Time Bar
During measurement, an
Elapsed time
bar will fill in as good samples of data are
processed. When 180 seconds of acceptable EEG samples have been acquired, the
bar will be filled in completely and the test will terminate automatically.
Troubleshooting
Refer to section 3.4 for troubleshooting suggestions.
Test Done Screen
At the end of a screening, the result of the most recently completed screening will
continue to be visible (Table 15). The ear selection control will return so that the user
can select the other ear for testing or can repeat a screening measure on the same
ear.
The ear selection control can also be used to toggle between the last ABR test
performed on the right and left ear during the session.
Table 15 Test Done Screen
Screen
Screen Area/
Graphic
Function Name Description
Figure 72
Patient
information
Shows current patient name
Ear selection
button returns
Choose ear to test
Test result bar,
result symbol
and other
measurement
information
Green check & bar = Pass
Black ? = Incomplete
Yellow X = Refer
Artifact = % “bad” samples
Time = duration of test in
“good” samples
Test list button
Shows the list of tests from this
session to select for printing
Start button
Begins next screening test on
this patient
Comment
Select to enter a comment
that saves with this test.
Back button
Returns to prior screen
5.11 OAE Test Screens
General
During the OAE process the screens will change to reflect the processes taking place
and the options available to the screener.
Summary of Contents for easyScreen