Operation Manual easyScreen
8508935 Rev. 4
22
04/04/2019
Figure 8
To access and clean the larger channel it is necessary to
remove the gasket from inside the probe tip. You can do
this using a fine pin. Push the gasket back into place after
cleaning (Figure 8).
Figure 9
Place the probe tip back onto the probe. Make sure that
the plastic pegs are inserted into the appropriate
corresponding cavities (Figure 9).
Figure 10
Screw the probe cap back on the probe. The force of
tightening the cap will tighten the screw sufficiently. Never
use tools to fix the probe cap! Use only your fingers and
screw it on until secure (Figure 10).
Figure 11
Only insert the cleaning floss or wire into the base of the
probe to ensure wax/debris is pushed out of the probe tip
instead of into it. This also protects the gasket from being
damaged (Figure 11).
If the inside of the probe, beyond the removable probe tip, is clogged or damaged, only
a MAICO authorized distributor can repair the parts. For further information ask your
authorized local distributor or MAICO.
Disposables
Use only the Sanibel Supply disposable supplies that are supplied with your
easyScreen system.
Eartips, EarCups
TM
and adhesive electrodes are intended
for single-use only. These should be discarded after use.
They cannot be cleaned.
In case of re-use of the single-use disposables, you
enhance the risk of cross-contamination!
Accessories/Replacement Parts
Some reusable components are subject to wear with use over time. We recommend
that you keep these replacement parts available (as appropriate for your easyScreen
device configuration).
WARNING