Magtrol Model DSP6001 Dynamometer Controller
Chapter 6 – Alarm System
SETUP • OPERA
TION
6.8
TEMPERATURE ALARM
• To alert user when dynamometer gets too hot and thermal switch opens
• Only available for use with WB or PB dynamometers
• Default - always active
6.8.1
i
nStrUCtionS
for
t
eMperatUre
a
LarM
S
etUp
No setup needed.
6.8.2
t
eMperatUre
a
LarM
a
Ction
When the dynamometer in use becomes too hot, the display will flash “TEMPERATURE ALARM
TSCX” (as indicated in
Figure 6–19
) and will be accompanied by an audible beeping sound.
The alarm relay will open and the excitation current will immediately decrease to 10%. After
approximately 3 seconds, the current wll drop to 0.
POWER
BRAKE STATUS
SET POINT
SET POINT
P
I
D
TORQUE
SPEED
STATUS
∗∗∗∗
∗∗∗∗
Figure 6–19 Temperature Alarm Message Display
6.8.3
t
o
r
eSet
t
eMperatUre
a
LarM
Press any front panel button other than SHIFT. The alarm condition must be cleared before the unit
will return to normal operation.