51
Magtrol Model DSP6001 Dynamometer Controller
Chapter 6 – Alarm System
SETUP • OPERA
TION
The relay contacts are made accessible on the Accessory Torque/Speed Output rear panel connector.
See
Figure 2–6.
COMMON
N.O.
N.C.
24VDC
Contactor coil
AC Mains
Motor
Under
Test
Figure 6–3 Typical Application
6.1.2
a
LarM
o
peration
The DSP6001 gives the user the ability to enable or disable the alarms in the unit. The default is set
in the OFF position. In order for the alarms to be operative the user must enable them.
6.1.2.1
How to Enable/Disable Alarms
1. Starting from main menu, press SHIFT.
2. Press DYNO SETUP button.
3. Select ALARMS.
4. Press SHIFT 3 times. The display should appear as follows:
POWER
BRAKE STATUS
SET POINT
SET POINT
P
I
D
TORQUE
SPEED
STATUS
Figure 6–4 Alarm Enable/Disable Menu
5. Press COM SETUP button until you reach desired selection (YES or NO).
6. Press SHIFT 2 times to return to the main menu.
Note:
Once the alarms are activated, they are only monitored for the
channel that the control loop is closed on. For example, if the
DSP6001 is controlling on TSC1, alarms for TSC2 are ignored.