Magtrol Model DSP6001 Dynamometer Controller
Chapter 6 – Alarm System
SETUP • OPERA
TION
6.1.3
a
LarM
p
riority
While in an alarm condition, a higher priority alarm will be acknowledged, while lower priority
alarms are ignored. The priority order is as follows.
Priority
Alarm
Availability
Hysteresis
Dynamometer
Eddy-Current/Powder
Brake Dynamometer
1
Electrical Alarm
N/A
X
2
Temperature Alarm
N/A
X
3
External Alarm
X
X
4
Air Flow Alarm
X
N/A
5
Water Flow Alarm
N/A
X
6
Maximum Torque
X
X
7
Maximum Speed
X
X
8
Power
X
X
6.2
POWER ALARM
• Used to indicate an over power condition
• Default is set at 1 kW
6.2.1
i
nStrUCtionS
for
p
ower
a
LarM
S
etUp
1. Starting from main menu, press SHIFT.
2. Press DYNO SETUP button.
3. Select MAXPOWER. The display should appear as follows:
POWER
BRAKE STATUS
SET POINT
SET POINT
P
I
D
TORQUE
SPEED
STATUS
Figure 6–5 Max Power Menu
4. Press MAX SPEED button and use UP
and DOWN
buttons and Decrease/Increase
dial to set desired maximum power value for TSC1.
5. Press SHIFT.
6. Press POWER UNITS button, then MAX SPEED button and use UP
and DOWN
buttons and Decrease/Increase dial to set desired maximum power value for TSC2.
7. Press SHIFT 3 times to complete Power Alarm setup and return to the main menu.