21
Magtrol AN 2000 C
Instruction Manual
The previously programmed INP30 value appears on the display, LED INP2 activated.
There are two methods to program input values:
Key-in method
: Use
to switch between "0" (positive) and "-" (negative).
Press
to advance to the next digit to the right, which goes in flash. Press
repeatedly
to increment the active digit until it takes desired value. Proceed in
the same manner with all the digits until desired value is completed on the display
with sign. Press
ENTER
to accept this value as INP30 and go next step.
Teach method
: Apply signal to the meter input. Press
TEACH
to view the actual
signal value present at the input connector, LED INP2 flashes. Press
ENTER
to accept
this value as INP30 and go next step.
ESC
To exit from the programming mode without saving changes.
Program the display value for the point 30, LED DSP30 activated. By means of the
and
procedure, program desired DSP30 value and press
ENTER
. The
limits of the span are -32000 and 32000 points. If the programmed value exceeds
from these limits, the meter indicates Error, then displays 32000 with the first digit in
flash to allow reprogramming the DSP12 value within limits.
ENTER
To save the entry into the memory and return to run mode.
ESC
Return to previous point.
1 second flag indication for scaling point 30.
[21.2] Input 30 value
B
A
TARE
HOLD
LIMIT
MAX
MIN
DATA
DSP2
INP2
FLT
STORE
DSP1
INP1
2
1
4
3
TARE
RESET
LIMIT
MAX/MIN
ENTER
ESC
DATA
PROG
TEACH
PRG
RUN
[21.3] Display 30 value
B
A
TARE
HOLD
LIMIT
MAX
MIN
DATA
DSP2
INP2
FLT
STORE
DSP1
INP1
2
1
4
3
TARE
RESET
LIMIT
MAX/MIN
ENTER
ESC
DATA
PROG
TEACH
PRG
RUN
[21.2] Point 30
B
A
HOLD
LIMI
MAX
MIN
DATA
DSP2
INP
STOR
DSP1
2
1
4
3
RESE
LIMI
MAX/M I
ENTE
ESC
DATA
PROG
TEAC
PR
RU