CALIBRATE TENSION USING MENU
www.maxcessintl.com
DLCA NET-ECAT
MI 850A365 1
Page 30 of 72
5.2
Weightless calibration
When editing a number, the digit being edited will flash. Use the
[
] button to select the digit to edit. Use the [+] and [-] buttons
to increment or decrement the digit number.
Enter the programming mode by pressing and holding [+] and [-]
for two seconds.
Use [+] and/or [-] to go to parameter P.02. Press [Enter] to see
the present load cell excitation voltage.
Press [Enter] to see the present load cell excitation voltage.
Select 7.5V if using MAGPOWR load cells.
Select either 5V or 7.5V if using non-MAGPOWR load cells
Press and hold [Enter] for two seconds (until the display starts
flashing) to store the changes.
Use [+] and/or [-] to go to CAL1.2 or CAL2.2. Press [Enter] to
enter the sub-menu.
Display now shows C12.1 or C22.2. Unload the load cell roller
and press [Enter]. The DLCA NET-ECAT will now zero out the idler
roll weight and show a wait display. After zeroing, the DLCA NET-
ECAT will automatically increment to the next parameter.
Display now shows C12.2 or C22.2. Press [Enter] and select the
decimal point position for tension. Press and hold [Enter] for two
seconds (until the display starts flashing) to store the decimal
point position. Range {000000, 000000., 00000.0, 0000.00,
000.000, 00.0000, 0.00000}.
Display now shows C12.3 or C22.3. Press [Enter] and enter the
load cell sensitivity in 0.1 mV/V units. MAGPOWR load cells have
a sensitivity of 2.1 mV/V. (Consult the manual of competitor’s
load cell if using non-MAGPOWR load cells.) Press and hold
[Enter] for two seconds (until the display starts flashing) to store
the parameter. Range {000.1, 100.0}.