How to Specify the Module Number and Counter Module ID
In the lists of commands,
①
indicates the module number and
②
indicates the counter module ID.
The module number in
①
is the setting value of the rotary switch on the front of the MG80-SC. If * is specified in
①
, the setting is applied regardless of the rotary switch setting value.
The counter module ID in
②
is the ID of the MG80-CM counter module connected to the stacking connector of the
MG80-SC main module. It is specified in hexadecimal format, from 0 to 15, in order starting from the main module.
Module number
Counter module ID
Meaning
0 to F
0 to F
Refers to the specified counter module ID of the
specified module number
0 to F
*
Refers to all counter modules with the specified
module number
*
*
Refers to all counter modules of the local
module number
*
0 to F
Refers to the specified counter module of the
local module number
In reference commands (commands ending in ?), if "*" is specified in
②
, no response will be made. It is
necessary to specify an individual counter module.
Note that commands for counter modules that do not exist are ignored.