Using Commands to Configure the Settings
7.5.1
Preparation
Configuration by command can be done via RS-232C or Ethernet connection. Prepare the following items.
No.
Product
Preparation
1
PC, PLC, etc.
Prepare a device that can send and receive commands via RS-232C or Ethernet.
RS-232C: Use serial communication terminal software.
Ethernet: Software that can send and receive commands via TCP/IP is required.
Send the commands to TCP/IP port 24000 on the MG80-SC.
2
Communication
cable
RS-232C: Use a DZ252 communication cable (sold separately).
Ethernet: LAN cable (CAT5e shielded type recommended)
7.5.2
Connect
1. Connect via RS-232C or Ethernet.
For Ethernet, set the IP address on the PC used for configuring the settings. Set the IP address of the PC so that it
is on the same network as the IP address of the MG80-SC.
Example: MG80-SC with IP address 192.168.0.100 and subnet 255.255.255.0
IP address : 192.168.0.xxx *
Subnet mask: 255.255.255.0
* Do not set xxx to the same address as another connected device.
2. Establish the connection as shown below.
LAN cable
CAT5e
Serial cable
DZ252
7.5.3
How to Configure the Settings
Use the setup commands that are listed in section 8.
If you use the PLC link setup commands described in section 8.6, use the Ethernet interface.