Magnescale MG80-SC1 Operating Manual Download Page 15

 

14

  / 

48

 

 

4.3

 

Removing the Unit from the DIN Rails 

While holding the unit in place so that it will not fall, pull down the DIN rail anchoring levers of all the modules until 

a clicking sound is heard. 

 

 

 

 

 

Summary of Contents for MG80-SC1

Page 1: ...MG80 SC1 MG80 SC2 Operating Manual RS 232C Ethernet Interface Module Read all the instructions in the manual carefully before use and strictly follow them Keep the manual for future references...

Page 2: ...N Rails 14 5 Specifications 15 6 Functions 16 6 1 Definition of Terms 16 6 2 Details of Functions 16 6 2 1 Peak hold 16 6 2 2 Preset 16 6 2 3 Reference Point 17 6 2 4 Reset 18 6 2 5 Start 18 6 2 6 Go...

Page 3: ...the Settings 38 8 Commands 39 8 1 Communication Data Output Format 39 8 2 How to Specify the Module Number and Counter Module ID 41 8 3 Setup Commands 42 8 4 Operation Commands 44 8 5 Read Commands 45...

Page 4: ...existing MG10 10A products MG10 10A users can continue to use their programming environment with just a few settings changes In addition the product has a PLC link function By establishing a PLC link...

Page 5: ...using this product the equipment to be used is listed below Name Description Image MG80 SC RS 232C interface unit Main module MG80 CM Counter module DK800S series Digital gauge DK805S DK812S DK830S DK...

Page 6: ...sing Command I F over RS 232C MG10 Replacement RS 232C Host PC or Digital gauge Up to 16 MG80 CMs MG80 SC Termination PLC 2 2 2 Connection Example Using Command I F over Ethernet TCP or UDP Host PC or...

Page 7: ...PLC Link RS 232C RS 232C Digital gauge Up to 16 MG80 CMs MG80 SC Termination PLC link compatible PLC 2 2 4 Connection Example over PLC Link Ethernet Digital gauge Up to 16 MG80 CMs MG80 SC Termination...

Page 8: ...atus of this module Color Status Meaning Green Unlit Power OFF Lit Measurement in progress with power ON Blinking Setup in progress with power ON ALARM lamp Lights red when any counter module within t...

Page 9: ...to match the terminal on the PLC side Compliant with EIA RS232C standard Signals Asynchronous start stop system half duplex system Communication speed 2400 9600 19200 38400 57600 115200 230400 bps Da...

Page 10: ...8pin male Dsub 9pin female Pin no Signal Signal GND Request to send Case Shield cable Case 8pin mini DIN 9pin Dsub Description Host unit side Receive data Transmit data Clear to send Connection diagra...

Page 11: ...10 48 DZ254 RS 232C cable Round 8pin male bare wire Pin no Cable color Signal 1 Red N C 2 White SG 0V 3 Black RXD 4 Yellow TXD 5 Blue CTS 6 Green RTS 7 Brown N C 8 Gray N C Case Shield...

Page 12: ...th 20 m or less I O connector This I O connector is used to operate all of the counter modules within the unit at once It has four input pins and one output pin MG80 SC1 Current sink type MG80 SC2 Cur...

Page 13: ...OFF 9600 bps OFF ON OFF 19200 bps OFF OFF ON 38400 bps OFF ON ON 57600 bps ON OFF OFF 115200 bps ON ON OFF 230400 bps ON OFF ON Italic typeface is used to indicate the factory settings If you are usi...

Page 14: ...cure the module 4 2 Installing the Unit on the DIN Rails This product supports 35 mm wide DIN rails When the interface unit is shipped from the factory the tabs on the DIN rail anchoring levers are lo...

Page 15: ...14 48 4 3 Removing the Unit from the DIN Rails While holding the unit in place so that it will not fall pull down the DIN rail anchoring levers of all the modules until a clicking sound is heard...

Page 16: ...et 100BASE TX 100 Mbit s Maximum cable length RS 232C 15 m 2 Ethernet 20 m CAT5e shielded type recommended 3 Module number setting By rotary switch Indicator lamps Power Green LED 1 Alarm Red LED 1 Ho...

Page 17: ...6 2 Details of Functions 6 2 1 Peak hold The MG80 SC holds the peak values at all times The peak values are checked by switching among the output value modes The start function makes it possible to r...

Page 18: ...there is a module that has been configured to use the reference point 2 Push the spindle of the measuring unit 1 mm or more and retract it When this is done for all the modules that have been configur...

Page 19: ...ue Sets the current value Peak to peak P P value Sets the value obtained by subtracting the minimum value zero from the maximum value 6 2 6 Go No Go Judgment Comparator This function judges go or no g...

Page 20: ...Latch In the current value mode this function holds output data and comparator output for that value When latched the current value will stop updating so the peak value will also stop updating Pause T...

Page 21: ...Pin Assignment No I O Function name Description 1 I Reset input All the measurement axes are reset 2 I RS 232C trigger input The measurement data of all the axes is latched and the data is output 3 I...

Page 22: ...ter data acquisition is finished Reset input start input Latch input pause input Data RS trigger input low speed mode RS trigger input high speed mode Input equivalent circuit common to MG80 SC1 SC2 O...

Page 23: ...ing communication protocols PLC manufacturer Connection method Communication protocol Mitsubishi Electric Corporation RS 232C MC protocol 1C frame format 4 RS 232C MC protocol 3C frame format 4 TCP UD...

Page 24: ...ations update parameter reset recall preset start and load reference point The operation sequence is as follows Each has a control flag and a corresponding completion flag When the PLC turns the contr...

Page 25: ...ator upper limit set 1 high order 0x0006 Comparator lower limit set 2 low order CL2 XXX 0x0007 Comparator lower limit set 2 high order 0x0008 Comparator upper limit set 2 low order CH2 XXX 0x0009 Comp...

Page 26: ...or 0 0 00001 mm 25 4 unit multiple of 2 1 m 1 m unit 0 0001 mm 25 4 unit 5 m 1 m unit last digit must be a 5 or 0 0 0001 mm 25 4 unit multiple of 2 10 m 10 m unit 1 0 001 mm 25 4 unit Measurement dat...

Page 27: ...Windows PC or by using commands Use whichever method is most convenient 7 2 Module Number Setting Use a screwdriver or other tool to align the direction of the arrow on the module number setting swit...

Page 28: ...pression software 7 4 2 Connect 1 Prepare the Setting application for Windows PC Place the downloaded MG80 SC_SettingTool exe file in the desired location 2 Set the IP address on the PC where Setting...

Page 29: ...application for Windows PC is installed When the program starts the following window will appear Follow the steps below to connect Enter the IP address of the MG80 SC to connect to If the module numbe...

Page 30: ...29 48 Once connected parameter setting is possible If the connection fails turn off the power of the MG80 SC exit the setting application and start over from the beginning...

Page 31: ...etting window Transitions to the PLC Link Setting window Transitions to the Measure Monitor window Indicates the module number of the connected MG80 SC Indicates the number of connected axes Specifies...

Page 32: ...direction input polarity of the measuring unit When this is selected the direction is reversed Specifies whether or not to use the reference point function When this is selected the reference point f...

Page 33: ...the following options External trigger low speed mode External trigger high speed mode Internal timer 0 2 seconds Internal timer 0 5 seconds Internal timer 1 0 seconds Internal timer 5 0 seconds Inte...

Page 34: ...sponding axis Enabled only when the reference point function is used If the reference point setting is changed the power must be turned off and on again Do not turn off the power of the MG80 SC while...

Page 35: ...rotocol to be used by the PLC link Check the available protocols for the PLC and select one of the following combinations PLC communication specifications MG80 SC settings PLC manufacturer Communicati...

Page 36: ...on PLC Data Memory DM Specifies the settings when FINS command is selected in SA1 Source FINS node address For UDP set the node address of the MG80 SC Specify the low order byte of the IP address of t...

Page 37: ...data If there is an error Error is displayed Indicates the measurement mode of the measured value REAL Current value MAX Maximum value MIN Minimum value P P Peak to peak value Indicates the comparator...

Page 38: ...About from the menu bar to display the Version Display screen Indicates the software name and version Closes the screen Enables the measurement unit switching function It is necessary to restart the...

Page 39: ...ld separately Ethernet LAN cable CAT5e shielded type recommended 7 5 2 Connect 1 Connect via RS 232C or Ethernet For Ethernet set the IP address on the PC used for configuring the settings Set the IP...

Page 40: ...L When an alarm has occurred E Numerical value Fixed at 8 bytes or sign followed by the data including decimal point Example 1 For a resolution of 0 1 m 99 9999 to 99 9999 Example 2 For a resolution...

Page 41: ...header output Mode 1 Normal data output Header consisting of 2 bytes module number counter module ID Example 00 09 9999 Mode 2 Data output with measuring mode information added Header consisting of 4...

Page 42: ...0 SC main module It is specified in hexadecimal format from 0 to 15 in order starting from the main module Module number Counter module ID Meaning 0 to F 0 to F Refers to the specified counter module...

Page 43: ...and Return format Preset value P The value differs depending on the measurement unit and input resolution Resolution mm unit mm 25 4 unit 0 1 m 0 5 m 99 9999 to 99 9999 99 9995 to 99 9995 9 99999 to 9...

Page 44: ...rn format Select the transfer data separator RSFORM 0 Space 1 Newline CR LF RSFORM RSFORM Command Return format Select the RS 232C output data format RSTRG 0 Mode 1 1 Mode 2 2 Mode 3 RSTRG RSTRG Comma...

Page 45: ...eration command it is not recorded in ROM CH CH CL CL Refer to the setup commands When executed as an operation command it is not recorded in ROM RES Command format Reset zero reset reference point se...

Page 46: ...and format Return format Load the reference point offset value LO LO The value differs depending on the measurement unit and input resolution VER Command format Return format Read the software version...

Page 47: ...n format PLC link use mode LMD 0 Do not use PLC link 1 Use PLC link LPC1 LPC1 Command Return format PLC link protocol 1 protocol LPC1 0 MC protocol 1C frame format 4 1 MC protocol 3C frame format 4 2...

Page 48: ...FINSDNA Command Return format FINS command parameter DNA FINSDNA 0 Own network 1 to 127 Destination network address Note Set this when using the FINS command FINSDA1 FINSDA1 Command Return format FIN...

Page 49: ...been inserted all the way in Are the same communication settings used for both the host unit and MG80 SC Is the cable too long Are any cables carrying heavy currents from motors or other devices posi...

Page 50: ...ent l usage des acqu reurs de l quipement d crit dans ce manuel Magnescale Co Ltd interdit formellement la copie de quelque partie que ce soit de ce manuel ou son emploi pour tout autre but que des op...

Page 51: ...kawa Isehara shi Kanagawa 259 1146 Japan 1 16 16 For foreign customers Note This product or technology may be restricted by the government in your country Please make sure that end use end user and co...