205
47DR OPERATING MANUAL EM-9509-B Rev.7
34.1.2 OPERATING PROCEDURE
NOTE
The following figures are display examples. The displays
depend on the settings.
1
Hold down Init/
↑
button for 3 seconds or more to move on to
Initial Setting Mode.
• The input type is indicated.
• The SD indicates ‘RTTYPE’.
2
Press Alarm/
↓
or Init/
↑
button to go to the input compensation
A input value setting.
• The input compensation A input value is indicated.
• The SD indicates ‘CR-A IN’.
3
Press Shift button to shift the display into the setting standby
mode. Press Shift and Up buttons to set the input compensa-
tion A input value.
• The sixth digit starts blinking, to which you can apply changes.
• Set within the range of -999.99 to 9999.99.
NOTE
Set to 2 decimal places regardless of the decimal point position setting.
4
Press Alarm/
↓
button to register the value and to go to the
input compensation A compensation value setting.
• The input compensation A input value is registered.
• The input compensation A compensation value is indicated.
• The SD indicates ‘CR-A CR’.
2,4,
6,8
1,2,
8
3,5
3,5
Blinking