Laser Console Basics 23
Novus Spectra
0642-543-01, Rev A, 12/02
Turning off the laser
Under normal operating conditions, turn the keyswitch to the
(off )
position to turn off the laser. Remove the key to prevent unauthorized use of
the laser.
Emergency off
In an emergency, press the
emergency off button on the front of the laser
console to immediately turn off the laser.
When the main power cable is connected to the electrical source, some
internal circuits remain energized. To de-energize all internal circuits, unplug the
main power plug from the wall socket, or turn off the main electrical service
(wall circuit breaker).
The emergency stop button turns all power to the laser off, but does not turn
off the CPU board or the laser console display. When the button is pressed, an
EOFF code will appear in the laser console display, and the system will be in
standby. To clear the EOFF, press the emergency stop button. The system will
return to standby mode with all of the user settings retained.