Novus Spectra
0642-543-01, Rev A, 12/02
30 General Operation
Setting Treatment Parameters
You can set treatment parameters from either the laser console display or the
remote control. The system beeps when a maximum available setting is
selected.
CAUTION -
Always verify that the desired treatment parameters are
displayed on the laser console and/or remote control before initiating
treatment. If there is no change in display values when you press the selectors
on the laser console, or if the laser console or remote control appears
otherwise erratic, do not use the laser. Contact your local Lumenis
representative.
Selecting and verifying the delivery device
Connect the desired delivery device, as described in “Connecting the delivery
device.” Verify that the appropriate delivery device icon illuminates on the
console display or remote control:
indicates a LaserLink or slit lamp is connected
indicates a probe is connected
indicates an LIO is connected
WARNING -
Always ensure that a compatible eye safety filter is properly
connected for the selected delivery device. Failure to properly use an eye
safety filter for those delivery devices that require them can result in serious
eye damage. Refer to the appropriate delivery device operator manual for
additional compatibility and safety information.
Adjusting the aiming beam intensity
Adjust the aiming beam intensity to the desired level by pressing the aiming
beam
(increase) and
(decrease) selectors. To adjust the aiming beam
intensity from the remote control, press the
(increase) and
(decrease)
selectors in the
(aiming beam) field.
During some procedures, the aiming beam may obstruct the view of the
treatment site while lasing. The aiming beam can be either on or off while
lasing. To turn the aiming beam off, press the down selector to the “0” setting.
Summary of Contents for Novus Spectra
Page 1: ...Novus Spectra Solid State 532 nm Green Laser Operator Manual...
Page 6: ...6 Contents Novus Spectra 0642 543 01 Rev A 12 02...
Page 8: ...Novus Spectra 0642 543 01 Rev A 12 02 8 General Operation...
Page 40: ...Novus Spectra 0642 543 01 Rev A 12 02 40 General Operation...
Page 42: ...Novus Spectra 0642 543 01 Rev A 12 02 42 Maintenance...
Page 56: ...Novus Spectra 0642 543 01 Rev A 12 02 56 Safety and Regulatory...
Page 72: ...Novus Spectra 0642 543 01 Rev A 12 02 72 Indications for Use...
Page 110: ...Novus Spectra 0642 543 01 Rev A 12 02 110 Indications for Use...