![LUMENIS Novus Spectra Operator'S Manual Download Page 19](http://html1.mh-extra.com/html/lumenis/novus-spectra/novus-spectra_operators-manual_1948101019.webp)
Connection Instructions
19
Novus Spectra
0642-543-01, Rev A, 12/02
Connecting eye safety filters (LaserLink, slit lamp, and Acculite only)
Installing the eye safety filter(s)
When using a LaserLink, a Lumenis slit lamp, or Acculite delivery devices,
you must install a compatible automatic eye safety filter or fixed filter onto
the slit lamp or operating microscope. The eye safety filter protects the
physician’s eyes from exposure to laser light while looking through the slit
lamp or operating microscope.
If the eye safety filter is not properly connected to the laser, an E400 error
code will display on the laser console. If the eye safety filter is not compatible,
an E401 error code will display.
Refer to the appropriate delivery device operator manual for detailed eye
safety filter installation instructions.
CAUTION -
During treatment, the eye safety filter must be attached to the
slit lamp or operating microscope. Failure to use an eye safety filter could
result in ocular injury. In addition, some slit lamps and microscopes require an
auxiliary eye safety filter to protect persons observing the procedure through
an observation tube. To determine if an auxiliary eye safety filter is required,
review your slit lamp or microscope operator manual or contact the
manufacturer.
Summary of Contents for Novus Spectra
Page 1: ...Novus Spectra Solid State 532 nm Green Laser Operator Manual...
Page 6: ...6 Contents Novus Spectra 0642 543 01 Rev A 12 02...
Page 8: ...Novus Spectra 0642 543 01 Rev A 12 02 8 General Operation...
Page 40: ...Novus Spectra 0642 543 01 Rev A 12 02 40 General Operation...
Page 42: ...Novus Spectra 0642 543 01 Rev A 12 02 42 Maintenance...
Page 56: ...Novus Spectra 0642 543 01 Rev A 12 02 56 Safety and Regulatory...
Page 72: ...Novus Spectra 0642 543 01 Rev A 12 02 72 Indications for Use...
Page 110: ...Novus Spectra 0642 543 01 Rev A 12 02 110 Indications for Use...