Novus Spectra
0642-543-01, Rev A, 12/02
46
Maintenance
F203
The footswitch has been depressed while
trying to enter ready mode.
Release the footswitch. Press the
ready button to continue, ensuring
that the footswitch is not depressed.
F204
Footswitch error
First check footswitch connections
and cord. If the condition persists,
contact your local Lumenis
representative.
F402
Delivery device error
First, remove and inspect the delivery
device as instructed in your delivery
device operator manual; then, verify
that the device is properly connected
as instructed. If the condition
persists, contact your local Lumenis
representative.
E501
The laser detected a lower level of energy
than the level selected by the user.
Turn off the system for five seconds,
then restart. If the condition persists,
contact your local Lumenis
representative.
E502
The laser detected a higher level of
energy than the level selected by the user.
Turn off the system for five seconds,
then restart. If the condition persists,
contact your local Lumenis
representative.
E701
The laser detected an over temperature
condition.
Turn off the system for 1 minute,
then restart. If the condition persists,
contact your local Lumenis
representative.
F803
Power interlock error
Turn the laser on and off. If the
condition persists, contact your local
Lumenis representative.
F503-F505, F601-F604,
F702-F704, F901-F909,
CA01
Internal system operating errors
Contact your local Lumenis
representative.
Problem or Error
Code
Probable Cause
Suggestion
Summary of Contents for Novus Spectra
Page 1: ...Novus Spectra Solid State 532 nm Green Laser Operator Manual...
Page 6: ...6 Contents Novus Spectra 0642 543 01 Rev A 12 02...
Page 8: ...Novus Spectra 0642 543 01 Rev A 12 02 8 General Operation...
Page 40: ...Novus Spectra 0642 543 01 Rev A 12 02 40 General Operation...
Page 42: ...Novus Spectra 0642 543 01 Rev A 12 02 42 Maintenance...
Page 56: ...Novus Spectra 0642 543 01 Rev A 12 02 56 Safety and Regulatory...
Page 72: ...Novus Spectra 0642 543 01 Rev A 12 02 72 Indications for Use...
Page 110: ...Novus Spectra 0642 543 01 Rev A 12 02 110 Indications for Use...