34 General Operation
Setting the treatment interval
The treatment interval is the time between laser pulses. Single and repeat
pulse modes are available. In single pulse mode, the laser delivers a single pulse
each time you depress the footswitch. In repeat pulse mode, the laser delivers
repeated pulses at a specified interval, from 0.05 s to 1.0 s, until you release
the footswitch.
Not all interval settings are available at higher power and exposure time
settings. If you attempt to select an interval that is unavailable at the existing
power and exposure time settings, the laser changes to meet the duty cycle
requirements for that power setting. You must reduce the power or exposure
time setting to make available the desired interval. Conversely, if you select the
minimum interval for the existing power and exposure time settings, and then
you increase the power or exposure time setting, the interval automatically
increases to the minimum setting for the new power and exposure time
settings.
To set the pulse interval, press the up or down Interval selectors on the laser
console display or the remote control until the desired value displays.
Summary of Contents for Novus Spectra
Page 1: ...Novus Spectra Solid State 532 nm Green Laser Operator Manual...
Page 6: ...6 Contents Novus Spectra 0642 543 01 Rev A 12 02...
Page 8: ...Novus Spectra 0642 543 01 Rev A 12 02 8 General Operation...
Page 40: ...Novus Spectra 0642 543 01 Rev A 12 02 40 General Operation...
Page 42: ...Novus Spectra 0642 543 01 Rev A 12 02 42 Maintenance...
Page 56: ...Novus Spectra 0642 543 01 Rev A 12 02 56 Safety and Regulatory...
Page 72: ...Novus Spectra 0642 543 01 Rev A 12 02 72 Indications for Use...
Page 110: ...Novus Spectra 0642 543 01 Rev A 12 02 110 Indications for Use...