ISR 12-LO/GS Manual
Data format UPP (Universal Pyrometer Protocol)
•
38
Description
Command
Parameters
Reading emissivity:
AAem
Output: XXXX (dec. 0010 ... 1000 in ‰)
Emissivity:
AAemXXXX
XXXX = (0010 ... 1000‰) (decimal)
Emissivity slope:
AAevXXXX
XXXX = 0800 ... 1200 (equivalent to 0.8 ... 1.2)
Ratio part in metal
mode:
AAmvXX *
)
XX = 01 ... 99%
Exposure time t
90
:
AAezX
X = 0 ... 6 (decimal) 0 = intrinsic time constant of the
device
1 = 0.01 s
3 = 0.25 s
5 = 3.00 s
2 = 0.05 s
4 = 1.00 s
6 = 10.00 s
External deletion:
AAlx
Simulation of an external deletion contact
Clear times of the
maximum value
storage:
AAlzX
X = 0 ... 9 (decimal) 0 = maximum value storage off
1 = 0.01 s
4 = 1.00 s
7 = external deletion
2 = 0.05 s
5 = 5.00 s
8 = automatically deletion
3 = 0.25 s
6 = 25.00 s
9 = Extern hold
Operation mode:
AAkaX
X = 0
metal mode from channel 1 and 2
X = 1
mono mode (1 channel)
X = 2 ratio temperature (2 channel)
Switch-of limit:
AAawXX
XX = 02 ... 50%
Analog output:
AAasX
X = 0 or 1 0 = 0 to 20 mA, 1 = 4 to 20 mA
Reading basic
temperature range:
AAmb
Output: XXXXYYYY (hex 8-digit, °C or °F)
XXXX = beginning of temperature range
YYYY = end of temperature range
Reading temperature
sub range:
AAme
Same as mb
Setting of temperature
sub range:
AAm1XXXXYYYY XXXX (hex 4-digit) beginning of temp. range (°C or °F)
YYYY (hex 4-digit) end of temp. range (°C or °F)
Address:
AAgaXX
XX = (00 ... 97)
00 ... 97 = regular device addresses
99 = global address with response
98 = global address without response (only setting
commands!)
Baud rate:
AAbrX
X = 1 ... 6 or 8 (dec.)
1 = 2400 baud
4 = 19200 baud
(7 is not available)
2 = 4800 baud
5 = 38400 baud
8 = 115200 baud
3 = 9600 baud
6 = 57600 baud
Temp. display °C / °F:
AAfhX
Output: X = 0: display in °C; X = 1: display in °F
Reading signal
strength:
AAtr
Output: XXXX, 1000 equivalent to 100.0%
Wait time:
AAtwXX
XX = 00 ... 99 (decimal)
Warning level
dirty window:
AAdwXX
XX = 00 … 99; 2 digit, dec. in %
Error status:
AAfs
Output:
XX;
XX=00…FF (00 = no error)
(01…FF: error code for LumaSense service)
Internal temperature:
AAgt
Output: XXX (dec. 000 ... 099, at °C, 032 ... 210 at °F)
Max. Internal temp.:
AAtm
Output: XXX (dec. 000 ... 099, at °C, 032 ... 210 at °F)
Reading interface type: AAin
Output: X = 1 or 2; 1 = RS232, 2 = RS485
Laser targeting light:
AAlaX
X = 0 switch off laser targeting light
X = 1 switch on laser targeting light