ISR 12-LO/GS Manual
Data format UPP (Universal Pyrometer Protocol)
•
39
Description
Command
Parameters
Lock keyboard:
AAlkX
X = 0 ... 3
1 = lock lk1, removal with command lk0 or power off-on
0 = removal of lock lk1
3 = continuous lock lk3, removal only with command lk2
2 = removal of lock lk3
Reading parameters:
AApa
Output decimal 11-digit:
Digit 1 und 2 (10...99 or 00):
ε
Digit 3 (0 ... 6): t
90
(exposition time)
Digit 4 (0 ... 8): t
CL
(max. storage clear mode)
Digit 5 (0 ... 1): analog output
Digit 6 und 7: (00 ... 98): temperature
Digit 8 und 9 (00 ... 97): address
Digit 10 (0 ... 6 or 8): baud rate
Digit 11 (0): (0 or 1): keyboard active or locked
Digit 12 ... 15 (0800 ... 1200): emissivity slope
Autoreset:
AAre
Activating auto-reset, initializes the pyrometer new and
resets the instrument to default settings
Reference number:
AAbn
Output: XXXXXX (hex 6-digit)
Serial number:
AAsn
Output: XXXX (hex 4-digit)
Device type /
software version:
AAve
Output: XXYYZZ (6-digit decimal)
XX = 06 (ISR 12-LO)
YY = month of software version
ZZ = year of software version
Software version
in detail:
AAvs
tt.mm.yy XX.YY
tt = day; mm = month; yy = year; XX.YY = software
version
Device type:
AAna
Output: “ISR 12-LO“ (16 ASCII-characters)
Special pouring stream commands
Description
Command
Parameters
Reading of data:
AAtg
Output: ABBBCCCC
A = serial no. of pouring date, Ascii-Hex 0 … F
BBB = time of pouring Ascii-Hex 000 … FFF in
1
/
10
s
(pre-run time + measuring time)
CCCC = temp. value Ascii-Hex 0000 … FFFF in
1
/
10
°C or °F
Pouring stream mode: AAtoXX
Output: XX XX = 00 pouring stream mode off
XX = 01 pouring stream mode on
Measuring condition:
AAtqXX
XX = 00 … 99; 2 digit, dec. in % of 1024 meas. values
Pre-run time:
AAtpXX
XX = 00 … 99; 2 digit, dec. in
1
/
10
s
Measuring time:
AAmtXX
XX = 00 … 99; 2 digit, dec. in
1
/
10
s; 00=Automatic mode
Follow-up time:
AAtnXX
XX = 00 … 99; 2 digit, dec. in
1
/
10
s
Note:
The letter “l” means the lower-case letter of “L”.